Philips CT Big Bore
by Philips Medical Systems (Cleveland), Inc. United States 510(k)
510(k) K171850
Product code JAK
Regulatory status
Recall ongoing. 5 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK171850 | Date9 Nov 2017 | StatusActive |
Coverage
- Active
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-2597-2025
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
-
- Ongoing Z-1660-2025
Unintended motion issues related to Interventional Control that may lead to collision of the Gantry/table with the operator or patient.
-
- Ongoing Z-0294-2024
- Ongoing Z-0293-2024
- Ongoing Z-0292-2024
For oncology users: If the user performed off-set reconstruction on CT device a shift could be observed on the contouring area of the primary and secondary images sets when using image fusion (on TPS) and incorrect coordinates may be sent to TPS system when using absolute patient marking. This may lead to incorrect radiation therapy planning and possible growth or spread of cancer due to incorrect early treatment regimen. For radiology users: If the user performed off-set reconstruction on CT devices, after multiple acquisitions, the Relate Position could be inaccurate. This could lead to incorrect diagnosis and treatment.
-
- Terminated Z-0309-2019
- Terminated Z-0311-2019
- Terminated Z-0307-2019
- Terminated Z-0310-2019
- Terminated Z-0303-2019
- Terminated Z-0304-2019
- Terminated Z-0306-2019
- Terminated Z-0308-2019
- Terminated Z-0305-2019
It was reported that scanners running certain software versions provide confusing information in the "Impressions" section of the Calcium Score report. The generated report provides multiple analyses of the amount of plaque detected, as if they all apply to the patient at the same time. This misleading guidance can be potentially misleading to the user, and may lead to mistreatment of the patient.
Specification
- 510(k) Number
- FDA Product Code
- JAK
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1750
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent