Philips Easy Diagnost
by Philips Medical Systems North America, Inc.
510(k) K970640
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK970640 | Date21 Mar 1997 | StatusActive |
Recalls
-
- Terminated Z-2208-2012
Inclination switch in the stand may have been mounted incorrectly, as a result, the switch could not stop the tilting movement of the table base due tol hardware or software errors.
-
- Terminated Z-1821-2012
If a lead ruler is not positioned in the same plane as the anatomical image detail to be measured, measured object distances (using the ruler) can be wrong by a corresponding factor. If, additionally, the measurement tool is not used to check the ruler, or is used in the right way (ie: calibration), such calibration error is not detected. As a result of this, a surgical intervention may be execut
-
- Terminated Z-0849-2010
Device failed to display appropriate labels indicating Radiation Symbol along with the required wording.
-
- Terminated Z-2405-2010
Potential for the tower (holding the serial changer and/or image intensifier) to slide unexpectedly and potentially injure a patient or user in its path.
-
- Terminated Z-0909-2007
Units do not comply with the requirement to provide automatic beam limitation within +/- 3 degrees of horizontal, vertical, and perpendicular to the plane of the image receptor during radiographic procedures.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- IXR
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1980
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent