Philips Easydiagnost Eleva
by Philips Medical Systems North America Co.
510(k) K031535
Active
Valid until
No expiry
FDA recalls
1 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK031535 | Date17 Jun 2003 | StatusActive |
Recalls
-
- Ongoing Z-0279-2024
Potential for units suspended on the ceiling to fail and fall.
-
- Terminated Z-2523-2020
Thermo switches in the main power supply for the system may be incorrectly installed, resulting in a malfunction.
-
- Terminated Z-0311-2018
If the lock mechanism of the movable camp of the mobile detector holder is not adjusted properly or not secured, the portable detector can drop down when the moveable clamp is at the bottom (below of the detector) and additional force is applied. This may happen when the detector is rotated clockwise from landscape to portrait position.
-
- Terminated Z-0365-2016
The hook does not securely hold the footplate in vertical position. Therefore the footplate can fall or swing down and may hurt the operator or patient.
-
- Terminated Z-0431-2014
The delivered application software was an outdated version to the model of equipment.
-
- Terminated Z-0767-2013
The mirror icon on the bottom of each image could be misinterpreted as a RIGHT lead marker for side indication if the image is mirrored within a PACS system.
-
- Terminated Z-2208-2012
Inclination switch in the stand may have been mounted incorrectly, as a result, the switch could not stop the tilting movement of the table base due tol hardware or software errors.
-
- Terminated Z-1821-2012
If a lead ruler is not positioned in the same plane as the anatomical image detail to be measured, measured object distances (using the ruler) can be wrong by a corresponding factor. If, additionally, the measurement tool is not used to check the ruler, or is used in the right way (ie: calibration), such calibration error is not detected. As a result of this, a surgical intervention may be execut
-
- Terminated Z-1283-2012
Underexposed image due to lag in switching between auxiliaries
-
- Terminated Z-1216-2012
Misinterpretation of Mirror Icon-Mirror icon on the bottom of each image could be misinterpreted as a RIGHT lead marker for side indication if the image is mirrored within a PACS system
-
- Terminated Z-0916-2011
Images from two different patients may be combined in one patient folder in the Digital Image (DI) or may be combined as one (printed) image. It is possible that an image of a previous patient registered in the DI will be overwritten with an image of a later patient.
-
- Terminated Z-0479-2011
Patient and image data can be exchanged when a patient is selected in the patient list and the examination tab is pressed with background query running at the same time. There is a risk that the data may then be linked to the incorrect patient, which may adversely affect diagnosis and treatment.
-
- Terminated Z-0353-2009
Tube failure occurs involving fluoroscopy of large patients and a partial failure of the grid switch.
-
- Terminated Z-2194-2010
Potential for bright artifacts to appear on the patient image from a previous exposure.
-
- Terminated Z-0910-2007
Units do not comply with the requirement to provide automatic beam limitation within +/- 3 degrees of horizontal, vertical, and perpendicular to the plane of the image receptor during radiographic procedures.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- KPR
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1680
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent
Other devices from Philips Medical Systems North America, Inc.
See all devices from Philips Medical Systems North America, Inc. →