Philips Integris Allura Flat Detector Release 1.2

by Philips Medical Systems North America, Inc.

510(k) K031333

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

2 ongoing

Recall on shared 510(k) K031333; other models may be affected

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK031333 Date13 May 2003ClearedStatusActive

Recalls

25 recalls · 2 ongoing · FDA

    • Ongoing Z-1077-2025 FDA Class II initiated 13 Jan 2025 on shared 510(k) K031333

    Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel

    • Ongoing Z-1151-2024 FDA Class II initiated 2 Jan 2024 on shared 510(k) K031333

    Disk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.

  • 2 recalls, same reason

    • Terminated Z-1140-2019 FDA Class II initiated 19 Feb 2019
    • Terminated Z-1141-2019 FDA Class II initiated 19 Feb 2019

    Additional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or continuous inability of making live fluoroscopy images or exposures.

  • 8 recalls, same reason

    • Terminated Z-3218-2018 FDA Class II initiated 16 Jul 2018
    • Terminated Z-3214-2018 FDA Class II initiated 16 Jul 2018
    • Terminated Z-3220-2018 FDA Class II initiated 16 Jul 2018
    • Terminated Z-3216-2018 FDA Class II initiated 16 Jul 2018
    • Terminated Z-3217-2018 FDA Class II initiated 16 Jul 2018
    • Terminated Z-3219-2018 FDA Class II initiated 16 Jul 2018
    • Terminated Z-3213-2018 FDA Class II initiated 16 Jul 2018
    • Terminated Z-3215-2018 FDA Class II initiated 16 Jul 2018

    There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.

  • 2 recalls, same reason

    • Terminated Z-1822-2017 FDA Class II initiated 22 Mar 2017 on shared 510(k) K031333
    • Terminated Z-1820-2017 FDA Class II initiated 22 Mar 2017 on shared 510(k) K031333

    Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.

    • Terminated Z-1280-2016 FDA Class II initiated 24 Sep 2014

    Loss of key image functionality due to a bent pedal of the Footswitch.

    • Terminated Z-1462-2012 FDA Not yet classified initiated 2 Apr 2012

    The monitor Ceiling suspension may fail, which may cause the monitor to drop.

    • Terminated Z-2976-2011 FDA Unclassified (pre-2012) initiated 8 Jul 2011

    System freezes from 20 seconds up to one minute. No fluoroscopy exposure and no geometry movements during this time period.

    • Terminated Z-2449-2010 FDA Unclassified (pre-2012) initiated 4 Feb 2010

    A lock-up might occur inside the Velara Generator due to intermittent shorts inside the system. The problem can only resolved by a warm or cold system restart. In exceptional cases the power converters of the KVmA unit in the generator are damaged and the system can not be restarted.

    • Terminated Z-0344-2009 FDA Unclassified (pre-2012) initiated 10 Dec 2008

    Tube failure occurs involving fluoroscopy of large patients and a partial failure of the grid switch.

    • Terminated Z-2471-2010 FDA Unclassified (pre-2012) initiated 23 Oct 2008

    A high voltage reliability issue resulting in the loss of the x-ray generation and this can cause an intermittent stop of fluoro acquisition and exposure during an examination without warning and hence no image on the display.

    • Terminated Z-0370-2007 FDA Unclassified (pre-2012) initiated 14 Dec 2006

    Movement of the AD5 table or Lateral C-arm (LARC). The table and/or LARC will not move upon request, will move slowly when movement is requested or will not stop moving when requested.

    • Terminated Z-0165-2007 FDA Unclassified (pre-2012) initiated 12 Oct 2006

    Potential for generator to lock-up. Fluoroscopy and x-ray is no longer possible, and system must be restarted to be able to continue.

    • Terminated Z-0581-06 FDA Unclassified (pre-2012) initiated 6 Feb 2006

    Potential lock-up of system requiring reboot of system.

  • 2 recalls, same reason

    • Terminated Z-0206-06 FDA Unclassified (pre-2012) initiated 12 Oct 2005
    • Terminated Z-0207-06 FDA Unclassified (pre-2012) initiated 12 Oct 2005

    Potential for high voltage exposure involving the injector interface connector of patient support tables.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
IZI
Device Class
Class II
Regulation
21 CFR Part 892.1600
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent