Philips Integris Allura Flat Detector Release 1.2
by Philips Medical Systems North America, Inc.
510(k) K031333
Active
Valid until
No expiry
FDA recalls
2 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK031333 | Date13 May 2003 | StatusActive |
Recalls
-
- Ongoing Z-1077-2025
Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
-
- Ongoing Z-1151-2024
Disk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
-
- Terminated Z-1140-2019
- Terminated Z-1141-2019
Additional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or continuous inability of making live fluoroscopy images or exposures.
-
- Terminated Z-3218-2018
- Terminated Z-3214-2018
- Terminated Z-3220-2018
- Terminated Z-3216-2018
- Terminated Z-3217-2018
- Terminated Z-3219-2018
- Terminated Z-3213-2018
- Terminated Z-3215-2018
There is a potential for the Monitor Ceiling Suspension, which holds a large screen monitor, may detach from the actuator rotor shaft causing the monitor to fall from the ground. When a Monitor Ceiling Suspension detaches from the actuator rotor shaft and the monitor falls, there is a risk of injury for the patient, user and bystander.
-
- Terminated Z-1822-2017
- Terminated Z-1820-2017
Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odor, smoke or fire.
-
- Terminated Z-1280-2016
Loss of key image functionality due to a bent pedal of the Footswitch.
-
- Terminated Z-1462-2012
The monitor Ceiling suspension may fail, which may cause the monitor to drop.
-
- Terminated Z-2976-2011
System freezes from 20 seconds up to one minute. No fluoroscopy exposure and no geometry movements during this time period.
-
- Terminated Z-2449-2010
A lock-up might occur inside the Velara Generator due to intermittent shorts inside the system. The problem can only resolved by a warm or cold system restart. In exceptional cases the power converters of the KVmA unit in the generator are damaged and the system can not be restarted.
-
- Terminated Z-0344-2009
Tube failure occurs involving fluoroscopy of large patients and a partial failure of the grid switch.
-
- Terminated Z-2471-2010
A high voltage reliability issue resulting in the loss of the x-ray generation and this can cause an intermittent stop of fluoro acquisition and exposure during an examination without warning and hence no image on the display.
-
- Terminated Z-0370-2007
Movement of the AD5 table or Lateral C-arm (LARC). The table and/or LARC will not move upon request, will move slowly when movement is requested or will not stop moving when requested.
-
- Terminated Z-0165-2007
Potential for generator to lock-up. Fluoroscopy and x-ray is no longer possible, and system must be restarted to be able to continue.
-
- Terminated Z-0581-06
Potential lock-up of system requiring reboot of system.
-
- Terminated Z-0206-06
- Terminated Z-0207-06
Potential for high voltage exposure involving the injector interface connector of patient support tables.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- IZI
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1600
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent