  "format": "devicerts.device.v1",
  "cite_as": "Philips Model Mx 16 Slice CT System, Philips Medical Systems (Cleveland), Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/philips-model-mx-16-slice-ct-system-jak (data as of 2026-11-22).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/philips-model-mx-16-slice-ct-system-jak",
    "json_url": "https://devicerts.com/devices/philips-model-mx-16-slice-ct-system-jak.json",
    "markdown_url": "https://devicerts.com/devices/philips-model-mx-16-slice-ct-system-jak.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-22",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/philips-model-mx-16-slice-ct-system-jak",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Philips Model Mx 16 Slice CT System",
    "registry_name": "PHILIPS MODEL MX 16 SLICE CT SYSTEM",
    "url": "https://devicerts.com/devices/philips-model-mx-16-slice-ct-system-jak",
    "manufacturer": {
      "name": "Philips Medical Systems (Cleveland), Inc.",
      "url": "https://devicerts.com/manufacturers/philips-medical-systems-cleveland-inc",
      "country": "United States",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K083498",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "JAK"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K083498",
      "number_label": null,
      "device_udi_di": null,
      "date": "2008-12-17",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-0190-2015",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2014-10-09",
      "reason": "\"The orientation of images is displayed incorrectly. When operator chooses gantry on the right side from operation console, and the patient position is HFDL, HFDR, FFDL or FFDR, the patient position picture and text is not consistent. \"L\" and \"R\" of scan images are reversed and this is not consistent with the patient actual position. \n\"When a surview which is longe",
      "applies": "this device",
      "via": "K083498"
    },
    {
      "number": "Z-0019-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2013-07-22",
      "reason": "A pinch point at the end of the table top can cause injury.",
      "applies": "this device",
      "via": "K083498"
    },
    {
      "number": "Z-0899-2013",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2012-12-19",
      "reason": "The four screws that secure the patient table top to the carrier became loose, which caused the table top to detach.",
      "applies": "this device",
      "via": "K083498"
    },
    {
      "number": "Z-2386-2012",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2012-08-24",
      "reason": "On August 24, 2012 the firm recalled the MX 16-slice after discovering that a bolt which attaches the oil accumulator within the gantry may fail, resulting in the oil accumulator detaching from the tube housing assembly within the gantry.",
      "applies": "this device",
      "via": "K083498"
    },
    {
      "number": "Z-2059-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-03-10",
      "reason": "The Philips MX 16-slice CT scanners are subject to recall/field correction due to a problem encountered with the units when using the 'Combine Viewing' Option.  The firm determined that inaccurate dimension measurements display when using the Combine Viewing Option. Moreover, if the dimension measurements from merged images created using this option are used, there is a potential for applying inac",
      "applies": "this device",
      "via": "K083498"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K083498",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "JAK",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 892.1750",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Radiology",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}