Philips Plus Computed Tomography System
by Philips Medical Systems, Inc.
510(k) K033326
Philips Plus Computed Tomography System is a medical device (FDA Class II; FDA product code JAK, System, X-Ray, Tomography, Computed) made by Philips Medical Systems, Inc., cleared under FDA 510(k) K033326 on 29 Oct 2003, active as of 22 Oct 2026. 97 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
6 ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK033326 | Date29 Oct 2003 | StatusActive |
Official records
- FDA 510(k) database K033326
From the FDA record
- Received by the FDA
- 16 Oct 2003
- FDA decision
- 29 Oct 2003 (Substantially Equivalent)
- FDA review time
- 13 days from receipt to decision
- Public document
- 510(k) summary on file with the FDA
Recalls
-
- Ongoing Z-3043-2026
Philips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment.
-
- Ongoing Z-2603-2025
- Ongoing Z-2604-2025
- Ongoing Z-2595-2025
- Ongoing Z-2602-2025
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
-
- Ongoing Z-1982-2021
Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.
-
- Terminated Z-0543-2019
- Terminated Z-0544-2019
There is a potential for separation of fiberglass resin which may cause the gantry front cover and the metal support on the cover to fail/fall
-
- Terminated Z-1143-2019
Philips has identified that the compensator within the X-ray tube collimator may fail, resulting in ring or smudge artifacts on images. The firm has identified systems that may be at risk for this failure due to the collimator compensator being past its lifetime expected usage.
-
- Terminated Z-2174-2018
The patient support head holder could contact the back ISO shelter wall during manual or motorized motion through the gantry bore. This may result in a serious injury that would require medical intervention to preclude permanent impairment in certain scenarios.
-
- Terminated Z-1396-2018
Surview scan with tube current lower than 30 mA is unable to be initialized.
-
- Terminated Z-1404-2018
- Terminated Z-1407-2018
- Terminated Z-1406-2018
- Terminated Z-1413-2018
- Terminated Z-1405-2018
Closing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.
-
- Terminated Z-0389-2018
Philips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the parts secured by the fasteners may become loose, which may have a potential for harm.
-
- Terminated Z-0694-2017
- Terminated Z-0693-2017
Software error due to the filament on timer
-
- Terminated Z-0011-2017
- Terminated Z-0010-2017
- Terminated Z-0012-2017
- Terminated Z-0009-2017
The infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may exhibit degraded image quality manifested as non-uniformity of the image, CT number shift greater than 5 Hounsfield units, and reduction in gray/white matter differentiation. The issue is most sever in the High Resolution mode, and the factory reference pediatric scan protocols are set to High Resolution default. Adults head scans have a factory reference protocol that defaults to standard resolution which doesn't experience the issue.
-
- Terminated Z-2652-2016
- Terminated Z-2658-2016
- Terminated Z-2657-2016
After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.
-
- Terminated Z-2114-2016
- Terminated Z-2115-2016
- Terminated Z-2113-2016
Supplier design change contributed to a thinner plastic which was used as part of the cover on the devices. The thinner plastic cover may crack causing injury to patients, operators, by-standers and service personnel.
-
- Terminated Z-2385-2016
- Terminated Z-2384-2016
Software issues in v4.1.3/4.1.4/4.1.5 in the Philips Ingenuity CT products that could affect the performance of the equipment.
-
- Terminated Z-2383-2016
Software issues in software versions v4.1.3/4.1.4/4.1.5 in the Philips Brilliance 64 and Ingenuity CT products that could affect the performance of the equipment.
-
- Terminated Z-2380-2016
Software issues found in software versions v4.1.3/4.1.4/4.1.5 in the Philips Brilliance 64 products that could affect the performance of the equipment.
-
- Terminated Z-1873-2016
- Terminated Z-1875-2016
- Terminated Z-1872-2016
- Terminated Z-1874-2016
The firm became aware of a problem where the system may not map Varian drive after CT user logout/login.
-
- Terminated Z-2116-2016
Philips Healthcare received reports from the field that certain Brilliance 64 systems running software version 3.5.4 exhibited intermittent swirl-like ring artifacts that may appear on reconstructed images. A patient rescan may be required if the images cannot be used for interpretation due to the swirl-like artifact.
-
- Terminated Z-2117-2016
Philips Healthcare received reports from the field that certain Ingenuity Core systems running software version 3.5.4 exhibited intermittent swirl-like ring artifacts that may appear on reconstructed images. A patient rescan may be required if the images cannot be used for interpretation due to the swirl-like artifact.
-
- Terminated Z-2118-2016
Philips Healthcare received reports from the field that certain Ingenuity Core128 systems running software version 3.5.4 exhibited intermittent swirl-like ring artifacts that may appear on reconstructed images. A patient rescan may be required if the images cannot be used for interpretation due to the swirl-like artifact.
-
- Terminated Z-2366-2016
- Terminated Z-2372-2016
- Terminated Z-2371-2016
Software defect causing intermittently slow response of Host.
-
- Terminated Z-1506-2016
The AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90 or 180 relative to one another. However, certain gamma cameras allow for other relative detector angles. Data acquired at these other angles will not be correctly reconstructed by AutoSPECT Pro and the results will likely be erroneous.
-
- Terminated Z-1351-2016
- Terminated Z-1356-2016
- Terminated Z-1354-2016
- Terminated Z-1355-2016
The firm was notified of a software error in which the system may not automatically send all image/data series to remote devices.
-
- Terminated Z-0547-2016
- Terminated Z-0549-2016
- Terminated Z-0548-2016
- Terminated Z-0544-2016
Perfusion scan feature may not be available on machines running software versions 4.1.2, 4.1.3, and 4.1.4. Customers previously using the basic axial perfusion software (versions preceding 4.1.2) without buying the helical perfusion license key could not access the basic axial perfusion feature after upgrading to software version 4.1.2.
-
- Terminated Z-0175-2016
- Terminated Z-0178-2016
- Terminated Z-0174-2016
- Terminated Z-0176-2016
The firm discovered Ring/Dot artifact due to X-ray measurement error.
-
- Terminated Z-1717-2016
- Terminated Z-1718-2016
- Terminated Z-1719-2016
- Terminated Z-1716-2016
Software Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of view; (4) DoseRight algorithm miscalculations leading to incorrect CTDI values; (5) scan length changes during subsequent axial results, (6) surview scan lengths near 135mm or 184mm
-
- Terminated Z-0406-2016
- Terminated Z-0408-2016
- Terminated Z-0405-2016
- Terminated Z-0407-2016
During scans with specific protocol steps, the software unexpectedly sets the Axial ( 2 axis ) scan length to a different length than the user set in the plan.
-
- Terminated Z-2205-2015
A head holder released for use with other CT systems but not yet qualified for use on the Ingenuity CT, which was shipped with this system.
-
- Terminated Z-2204-2015
A head holder released for use with other CT systems but not yet qualified for use on the Ingenuity Core 128, which was shipped with this system.
-
- Terminated Z-2203-2015
A head holder released for use with other CT systems but not yet qualified for use on the Ingenuity Core, which was shipped with this system.
-
- Terminated Z-1657-2015
- Terminated Z-1649-2015
- Terminated Z-1650-2015
The firm was informed that while raising the patient couch on the system to perform an exam, the couch unexpectedly descended to the lowest point without being commanded to do so.
-
- Terminated Z-1773-2015
- Terminated Z-1772-2015
- Terminated Z-1761-2015
- Terminated Z-1770-2015
- Terminated Z-1771-2015
- Terminated Z-1762-2015
The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.
-
- Terminated Z-2259-2015
- Terminated Z-2258-2015
- Terminated Z-2257-2015
- Terminated Z-2254-2015
5 issues. Fast Get Ready incorrectly enabled during service procedures. Contrast annotation missing for some images when manual contrast was administered. Dot artifact present, intermittently, after startup. Heart Rate measured value incorrectly remains constant for some heart rate change conditions. DoseRight incorrectly enabled when the Exam Card specified disabling DoseRight. Tube heat predict
-
- Terminated Z-1359-2015
Philips discovered that a software defect exists in marketed product wherein the sign indication of the longitudinal position of some types of scan is inverted.
-
- Terminated Z-0979-2015
Improper assembly of the vertical motor/brake system may lead to uncommanded vertical motion.
-
- Terminated Z-2711-2014
Ingenuity CT system can mislabel the exposure that results in an image in the incorrect order that it was taken by the user.
-
- Terminated Z-2237-2014
The following software issues have been identified in the affected products. Problem 1: When reopening a bookmark generated from processing a MUGA (Multi-Gated Acquisition) scan within the NM Cardiac Application, the ejection fraction (EF) displayed should be the same as the ejection fraction (EF) originally displayed when the bookmark was first created. In some instances, the ejection fraction (
-
- Terminated Z-1722-2014
A customer reported that the table top had become free floating. A Field Service Engineer (FSE) confirmed that the service latch had become disengaged.
-
- Terminated Z-1461-2014
Block assembly-lock stop subframe of service latch was not manufactured to engineering specifications from the supplier, causing premature fracture of the threaded rod on the patient support service latch. As a result, the table top can become free floating due to the disengaged service latch in the table tops subframe.
-
- Terminated Z-1678-2014
During SPECT reconstructions using Attenuation Correction and Scatter Correction, no Scatter Correction is being applied in the AutoSPECT Pro application on IntelliSpace Portal. In addition, resolution recovery is not applied correctly in SPECT reconstructions using Astonish in AutoSPECT Pro on Intellispace Portal.
-
- Terminated Z-1464-2014
A customer reported that after upgrading to software version 3.5.5 from 3.5.4, reconstructions would intermittently not complete.
-
- Terminated Z-2210-2013
The patient support may move in an unintended manner if the footswitch cover impinges on the footswitch.
-
- Terminated Z-1622-2013
MAQUET has identified a potential issue with the extension device handle (part number 1433.62A1) and will make enhancements to the YUNO table (part number 1433.01F0). Identification of the Extension device issue: By fully actuating the lever of the extension device handle, the operator's skin can enter a gap between the handle lever and a recess in the handle bar. When the lever is released,
-
- Terminated Z-1646-2013
Philips Healthcare received reports from the field that certain Brilliance CT 64, Ingenuity Core, and Ingenuity Core128 systems running software versions 2.6.1 or 3.5.4 exhibited intermittent swirl-like ring artifacts that may appear on reconstructed images, which may impair the diagnostic quality of affected images.
-
- Terminated Z-1588-2013
If the operator selects a patient name from the worklist using the mouse and then selects a different patient name using the keyboard (arrow keys then press enter) the patient name that populates the exam information page is the patient name selected with the mouse. Also, it is possible to begin a normal clinical scan while the anti-virus scan is running. However, if the user is running a bol
-
- Terminated Z-0950-2013
This field change order is being released to update software and customer release notes to the affected installed base.
-
- Terminated Z-0945-2013
Philips Healthcare received reports from the field stating the patient table on a system had an unexpected downward movement while a patient was on the table. Philips has determined the cause was a malfunction of the vertical brake.
-
- Terminated Z-0674-2013
Software anomalies. Philips Medical Systems recalled their Brilliance 64 running software v4.0.0xx516 and their Ingenuity CT running software v4.0.0xx518 due to software issues detected in these CT scanning systems. If these issues were to re-occur they could pose the risk of injury to patients undergoing scans; device users; or device service personnel.
-
- Terminated Z-2245-2012
The patient support may move in an unintended manner if the foot switch cover becomes damaged and impinges on the foot switch.
-
- Terminated Z-2223-2012
1. The current system design of software version 4.0.0xx379 does not prevent the system from scanning if the dose level exceeds a set upper dose limit for CT Brain Perfusion scanning, which previous versions did. However, the current version does have the Dose Check (NEMA XR-25) feature, which provides dose notifications and alerts, but it is possible for the user to disable Dose Check after ente
-
- Terminated Z-2035-2012
Philips Healthcare was notified that several of the sixteen bolts securing the brackets which support the carbon table top on various models of their Brilliant CT scanner have been found to be defective and the bolt heads may break off.
-
- Terminated Z-1988-2012
Philips is issuing a software update to the Intellispace Portal systems to address a number of non-conformances. The most frequent and severe non-conformance identified involves the portal setting of "PAC Integration". This setting has a feature that allows the user to launch images of multiple patient studies into an application, but does not display a warning message informing the user that diff
-
- Terminated Z-1987-2012
The Intellispace Portal software and Extended Brillance Workspace software may have significantly different results in the colors and size of the affected area when the customer uses the "Time Sensitive" and "Time Insensitive" methods.
Timeline
- Recall initiated FDA recalls Z-2174-2018
- Recall initiated FDA recalls Z-1143-2019
- Recall initiated FDA recalls Z-0543-2019
- Recall initiated FDA recalls Z-0544-2019
- Recall initiated FDA recalls Z-1982-2021
- Recall initiated FDA recalls Z-2603-2025
- Recall initiated FDA recalls Z-2604-2025
- Recall initiated FDA recalls Z-2595-2025
- Recall initiated FDA recalls Z-2602-2025
- Recall initiated FDA recalls Z-3043-2026
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- JAK
- Device class
- Class II
- Regulation
- 21 CFR Part 892.1750
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code JAK (System, X-Ray, Tomography, Computed).
- SYNGO NEURO PBV CT by Siemens AG
- CT SCANNER DIAGNOSTIC CONSOLE, REMOTE by Varian Assoc., Inc.
- GE LIGHTSPEED 7.2 CT SCANNER SYSTEM by GE Medical Systems, LLC
- LUNG VOLUME ANALYSIS by Toshiba Medical Systems
- CT BIOPSYVIEW by Philips Medical Systems North America, Inc.
- INSTANT IMAGER by Schiff Photo Mechanics
- ULTRAGUIDE CTG 2000SA by Ultraguide, Ltd.
- PercuNav Image Fusion and Interventional Navigation by Philips Ultrasound LLC
- Aquilion Prime SP, TSX-303B/1, v8.4 by Toshiba Medical Systems Coroporation
- ADVANTAGE WINDOWS VOLUME RENDERING OPTION by GE MEDICAL SYSTEMS, INC.