Philips Plus Computed Tomography System

by Philips Medical Systems, Inc.

510(k) K033326

Philips Plus Computed Tomography System is a medical device (FDA Class II; FDA product code JAK, System, X-Ray, Tomography, Computed) made by Philips Medical Systems, Inc., cleared under FDA 510(k) K033326 on 29 Oct 2003, active as of 22 Oct 2026. 97 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

6 ongoing

This device is named in the recall

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK033326 Date29 Oct 2003ClearedStatusActive

Official records

From the FDA record

Received by the FDA
16 Oct 2003
FDA decision
29 Oct 2003 (Substantially Equivalent)
FDA review time
13 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K033326 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

100 recalls · 6 ongoing · FDA

    • Ongoing Z-3043-2026 FDA Class II initiated 7 Aug 2026

    Philips has identified an unintended radiation issue with the Philips IQon, iCT, Ingenuity, and Brilliance CT 64 systems that could affect the performance of the equipment.

  • 4 recalls, same reason

    • Ongoing Z-2603-2025 FDA Class II initiated 13 Aug 2025
    • Ongoing Z-2604-2025 FDA Class II initiated 13 Aug 2025
    • Ongoing Z-2595-2025 FDA Class II initiated 13 Aug 2025
    • Ongoing Z-2602-2025 FDA Class II initiated 13 Aug 2025

    The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.

    • Ongoing Z-1982-2021 FDA Class II initiated 1 May 2021

    Significant wear of the Monitor Bracket Assembly shaft goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the cables resulting in patient, operator, bystander, or service person to serious injury.

  • 2 recalls, same reason

    • Terminated Z-0543-2019 FDA Class II initiated 18 Oct 2018
    • Terminated Z-0544-2019 FDA Class II initiated 18 Oct 2018

    There is a potential for separation of fiberglass resin which may cause the gantry front cover and the metal support on the cover to fail/fall

    • Terminated Z-1143-2019 FDA Class II initiated 3 Jul 2018

    Philips has identified that the compensator within the X-ray tube collimator may fail, resulting in ring or smudge artifacts on images. The firm has identified systems that may be at risk for this failure due to the collimator compensator being past its lifetime expected usage.

    • Terminated Z-2174-2018 FDA Class II initiated 30 Mar 2018

    The patient support head holder could contact the back ISO shelter wall during manual or motorized motion through the gantry bore. This may result in a serious injury that would require medical intervention to preclude permanent impairment in certain scenarios.

    • Terminated Z-1396-2018 FDA Class II initiated 21 Feb 2018

    Surview scan with tube current lower than 30 mA is unable to be initialized.

  • 5 recalls, same reason

    • Terminated Z-1404-2018 FDA Class II initiated 8 Feb 2018
    • Terminated Z-1407-2018 FDA Class II initiated 8 Feb 2018
    • Terminated Z-1406-2018 FDA Class II initiated 8 Feb 2018
    • Terminated Z-1413-2018 FDA Class II initiated 8 Feb 2018
    • Terminated Z-1405-2018 FDA Class II initiated 8 Feb 2018

    Closing the Emergency Stop may disable the horizontal motor, allowing the couch top to move during patient load or unload activities.

    • Terminated Z-0389-2018 FDA Class II initiated 10 Nov 2017

    Philips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the parts secured by the fasteners may become loose, which may have a potential for harm.

  • 2 recalls, same reason

    • Terminated Z-0694-2017 FDA Class II initiated 12 Nov 2016
    • Terminated Z-0693-2017 FDA Class II initiated 12 Nov 2016

    Software error due to the filament on timer

  • 4 recalls, same reason

    • Terminated Z-0011-2017 FDA Class II initiated 24 Aug 2016
    • Terminated Z-0010-2017 FDA Class II initiated 24 Aug 2016
    • Terminated Z-0012-2017 FDA Class II initiated 24 Aug 2016
    • Terminated Z-0009-2017 FDA Class II initiated 24 Aug 2016

    The infant performance test did not meet the test specification for the Infant Head Uniformity test with High Resolution. As a result, when the user performs High Resolution head scans, the reconstructed images may exhibit degraded image quality manifested as non-uniformity of the image, CT number shift greater than 5 Hounsfield units, and reduction in gray/white matter differentiation. The issue is most sever in the High Resolution mode, and the factory reference pediatric scan protocols are set to High Resolution default. Adults head scans have a factory reference protocol that defaults to standard resolution which doesn't experience the issue.

  • 3 recalls, same reason

    • Terminated Z-2652-2016 FDA Class II initiated 20 Jul 2016
    • Terminated Z-2658-2016 FDA Class II initiated 20 Jul 2016
    • Terminated Z-2657-2016 FDA Class II initiated 20 Jul 2016

    After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.

  • 3 recalls, same reason

    • Terminated Z-2114-2016 FDA Class II initiated 21 Jun 2016
    • Terminated Z-2115-2016 FDA Class II initiated 21 Jun 2016
    • Terminated Z-2113-2016 FDA Class II initiated 21 Jun 2016

    Supplier design change contributed to a thinner plastic which was used as part of the cover on the devices. The thinner plastic cover may crack causing injury to patients, operators, by-standers and service personnel.

  • 2 recalls, same reason

    • Terminated Z-2385-2016 FDA Class II initiated 29 Apr 2016
    • Terminated Z-2384-2016 FDA Class II initiated 29 Apr 2016

    Software issues in v4.1.3/4.1.4/4.1.5 in the Philips Ingenuity CT products that could affect the performance of the equipment.

    • Terminated Z-2383-2016 FDA Class II initiated 29 Apr 2016

    Software issues in software versions v4.1.3/4.1.4/4.1.5 in the Philips Brilliance 64 and Ingenuity CT products that could affect the performance of the equipment.

    • Terminated Z-2380-2016 FDA Class II initiated 29 Apr 2016

    Software issues found in software versions v4.1.3/4.1.4/4.1.5 in the Philips Brilliance 64 products that could affect the performance of the equipment.

  • 4 recalls, same reason

    • Terminated Z-1873-2016 FDA Class II initiated 8 Apr 2016
    • Terminated Z-1875-2016 FDA Class II initiated 8 Apr 2016
    • Terminated Z-1872-2016 FDA Class II initiated 8 Apr 2016
    • Terminated Z-1874-2016 FDA Class II initiated 8 Apr 2016

    The firm became aware of a problem where the system may not map Varian drive after CT user logout/login.

    • Terminated Z-2116-2016 FDA Class II initiated 1 Apr 2016

    Philips Healthcare received reports from the field that certain Brilliance 64 systems running software version 3.5.4 exhibited intermittent swirl-like ring artifacts that may appear on reconstructed images. A patient rescan may be required if the images cannot be used for interpretation due to the swirl-like artifact.

    • Terminated Z-2117-2016 FDA Class II initiated 1 Apr 2016

    Philips Healthcare received reports from the field that certain Ingenuity Core systems running software version 3.5.4 exhibited intermittent swirl-like ring artifacts that may appear on reconstructed images. A patient rescan may be required if the images cannot be used for interpretation due to the swirl-like artifact.

    • Terminated Z-2118-2016 FDA Class II initiated 1 Apr 2016

    Philips Healthcare received reports from the field that certain Ingenuity Core128 systems running software version 3.5.4 exhibited intermittent swirl-like ring artifacts that may appear on reconstructed images. A patient rescan may be required if the images cannot be used for interpretation due to the swirl-like artifact.

  • 3 recalls, same reason

    • Terminated Z-2366-2016 FDA Class II initiated 24 Mar 2016
    • Terminated Z-2372-2016 FDA Class II initiated 24 Mar 2016
    • Terminated Z-2371-2016 FDA Class II initiated 24 Mar 2016

    Software defect causing intermittently slow response of Host.

    • Terminated Z-1506-2016 FDA Class II initiated 4 Mar 2016

    The AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90 or 180 relative to one another. However, certain gamma cameras allow for other relative detector angles. Data acquired at these other angles will not be correctly reconstructed by AutoSPECT Pro and the results will likely be erroneous.

  • 4 recalls, same reason

    • Terminated Z-1351-2016 FDA Class II initiated 5 Feb 2016
    • Terminated Z-1356-2016 FDA Class II initiated 5 Feb 2016
    • Terminated Z-1354-2016 FDA Class II initiated 5 Feb 2016
    • Terminated Z-1355-2016 FDA Class II initiated 5 Feb 2016

    The firm was notified of a software error in which the system may not automatically send all image/data series to remote devices.

  • 4 recalls, same reason

    • Terminated Z-0547-2016 FDA Class II initiated 29 Oct 2015
    • Terminated Z-0549-2016 FDA Class II initiated 29 Oct 2015
    • Terminated Z-0548-2016 FDA Class II initiated 29 Oct 2015
    • Terminated Z-0544-2016 FDA Class II initiated 29 Oct 2015

    Perfusion scan feature may not be available on machines running software versions 4.1.2, 4.1.3, and 4.1.4. Customers previously using the basic axial perfusion software (versions preceding 4.1.2) without buying the helical perfusion license key could not access the basic axial perfusion feature after upgrading to software version 4.1.2.

  • 4 recalls, same reason

    • Terminated Z-0175-2016 FDA Class II initiated 14 Sep 2015
    • Terminated Z-0178-2016 FDA Class II initiated 14 Sep 2015
    • Terminated Z-0174-2016 FDA Class II initiated 14 Sep 2015
    • Terminated Z-0176-2016 FDA Class II initiated 14 Sep 2015

    The firm discovered Ring/Dot artifact due to X-ray measurement error.

  • 4 recalls, same reason

    • Terminated Z-1717-2016 FDA Class II initiated 6 Sep 2015
    • Terminated Z-1718-2016 FDA Class II initiated 6 Sep 2015
    • Terminated Z-1719-2016 FDA Class II initiated 6 Sep 2015
    • Terminated Z-1716-2016 FDA Class II initiated 6 Sep 2015

    Software Defects resulting in: (1) sagittal result shortened for axial scans; (2) single series displayed on console viewer when Creating multi-planner reconstruction; (3) scan length changes with a change in field of view; (4) DoseRight algorithm miscalculations leading to incorrect CTDI values; (5) scan length changes during subsequent axial results, (6) surview scan lengths near 135mm or 184mm

  • 4 recalls, same reason

    • Terminated Z-0406-2016 FDA Class II initiated 14 Apr 2015
    • Terminated Z-0408-2016 FDA Class II initiated 14 Apr 2015
    • Terminated Z-0405-2016 FDA Class II initiated 14 Apr 2015
    • Terminated Z-0407-2016 FDA Class II initiated 14 Apr 2015

    During scans with specific protocol steps, the software unexpectedly sets the Axial ( 2 axis ) scan length to a different length than the user set in the plan.

    • Terminated Z-2205-2015 FDA Class II initiated 17 Mar 2015

    A head holder released for use with other CT systems but not yet qualified for use on the Ingenuity CT, which was shipped with this system.

    • Terminated Z-2204-2015 FDA Class II initiated 17 Mar 2015

    A head holder released for use with other CT systems but not yet qualified for use on the Ingenuity Core 128, which was shipped with this system.

    • Terminated Z-2203-2015 FDA Class II initiated 17 Mar 2015

    A head holder released for use with other CT systems but not yet qualified for use on the Ingenuity Core, which was shipped with this system.

  • 3 recalls, same reason

    • Terminated Z-1657-2015 FDA Class II initiated 16 Mar 2015
    • Terminated Z-1649-2015 FDA Class II initiated 16 Mar 2015
    • Terminated Z-1650-2015 FDA Class II initiated 16 Mar 2015

    The firm was informed that while raising the patient couch on the system to perform an exam, the couch unexpectedly descended to the lowest point without being commanded to do so.

  • 6 recalls, same reason

    • Terminated Z-1773-2015 FDA Class II initiated 6 Mar 2015
    • Terminated Z-1772-2015 FDA Class II initiated 6 Mar 2015
    • Terminated Z-1761-2015 FDA Class II initiated 6 Mar 2015
    • Terminated Z-1770-2015 FDA Class II initiated 6 Mar 2015
    • Terminated Z-1771-2015 FDA Class II initiated 6 Mar 2015
    • Terminated Z-1762-2015 FDA Class II initiated 6 Mar 2015

    The firm was notified that the failure of the UPS accessory devices used in conjunction with Computed Tomography X-ray systems and Diagnostic Imaging systems. may result in battery acid leakage, overheating and/or emission of fumes.

  • 4 recalls, same reason

    • Terminated Z-2259-2015 FDA Class II initiated 5 Feb 2015
    • Terminated Z-2258-2015 FDA Class II initiated 5 Feb 2015
    • Terminated Z-2257-2015 FDA Class II initiated 5 Feb 2015
    • Terminated Z-2254-2015 FDA Class II initiated 5 Feb 2015

    5 issues. Fast Get Ready incorrectly enabled during service procedures. Contrast annotation missing for some images when manual contrast was administered. Dot artifact present, intermittently, after startup. Heart Rate measured value incorrectly remains constant for some heart rate change conditions. DoseRight incorrectly enabled when the Exam Card specified disabling DoseRight. Tube heat predict

    • Terminated Z-1359-2015 FDA Class II initiated 5 Feb 2015

    Philips discovered that a software defect exists in marketed product wherein the sign indication of the longitudinal position of some types of scan is inverted.

    • Terminated Z-0979-2015 FDA Class II initiated 8 Dec 2014

    Improper assembly of the vertical motor/brake system may lead to uncommanded vertical motion.

    • Terminated Z-2711-2014 FDA Class II initiated 11 Sep 2014

    Ingenuity CT system can mislabel the exposure that results in an image in the incorrect order that it was taken by the user.

    • Terminated Z-2237-2014 FDA Class II initiated 27 Jul 2014

    The following software issues have been identified in the affected products. Problem 1: When reopening a bookmark generated from processing a MUGA (Multi-Gated Acquisition) scan within the NM Cardiac Application, the ejection fraction (EF) displayed should be the same as the ejection fraction (EF) originally displayed when the bookmark was first created. In some instances, the ejection fraction (

    • Terminated Z-1722-2014 FDA Class II initiated 1 Apr 2014

    A customer reported that the table top had become free floating. A Field Service Engineer (FSE) confirmed that the service latch had become disengaged.

    • Terminated Z-1461-2014 FDA Class II initiated 27 Mar 2014

    Block assembly-lock stop subframe of service latch was not manufactured to engineering specifications from the supplier, causing premature fracture of the threaded rod on the patient support service latch. As a result, the table top can become free floating due to the disengaged service latch in the table tops subframe.

    • Terminated Z-1678-2014 FDA Class II initiated 11 Mar 2014 on shared 510(k) K033326

    During SPECT reconstructions using Attenuation Correction and Scatter Correction, no Scatter Correction is being applied in the AutoSPECT Pro application on IntelliSpace Portal. In addition, resolution recovery is not applied correctly in SPECT reconstructions using Astonish in AutoSPECT Pro on Intellispace Portal.

    • Terminated Z-1464-2014 FDA Class II initiated 6 Mar 2014 on shared 510(k) K033326

    A customer reported that after upgrading to software version 3.5.5 from 3.5.4, reconstructions would intermittently not complete.

    • Terminated Z-2210-2013 FDA Class II initiated 5 Jul 2013

    The patient support may move in an unintended manner if the footswitch cover impinges on the footswitch.

    • Terminated Z-1622-2013 FDA Class II initiated 26 Apr 2013 on shared 510(k) K033326

    MAQUET has identified a potential issue with the extension device handle (part number 1433.62A1) and will make enhancements to the YUNO table (part number 1433.01F0). Identification of the Extension device issue: By fully actuating the lever of the extension device handle, the operator's skin can enter a gap between the handle lever and a recess in the handle bar. When the lever is released,

    • Terminated Z-1646-2013 FDA Class II initiated 15 Apr 2013

    Philips Healthcare received reports from the field that certain Brilliance CT 64, Ingenuity Core, and Ingenuity Core128 systems running software versions 2.6.1 or 3.5.4 exhibited intermittent swirl-like ring artifacts that may appear on reconstructed images, which may impair the diagnostic quality of affected images.

    • Terminated Z-1588-2013 FDA Class II initiated 21 Mar 2013

    If the operator selects a patient name from the worklist using the mouse and then selects a different patient name using the keyboard (arrow keys then press enter) the patient name that populates the exam information page is the patient name selected with the mouse. Also, it is possible to begin a normal clinical scan while the anti-virus scan is running. However, if the user is running a bol

    • Terminated Z-0950-2013 FDA Class II initiated 15 Feb 2013

    This field change order is being released to update software and customer release notes to the affected installed base.

    • Terminated Z-0945-2013 FDA Class II initiated 20 Dec 2012

    Philips Healthcare received reports from the field stating the patient table on a system had an unexpected downward movement while a patient was on the table. Philips has determined the cause was a malfunction of the vertical brake.

    • Terminated Z-0674-2013 FDA Class II initiated 3 Dec 2012

    Software anomalies. Philips Medical Systems recalled their Brilliance 64 running software v4.0.0xx516 and their Ingenuity CT running software v4.0.0xx518 due to software issues detected in these CT scanning systems. If these issues were to re-occur they could pose the risk of injury to patients undergoing scans; device users; or device service personnel.

    • Terminated Z-2245-2012 FDA Class II initiated 3 Aug 2012

    The patient support may move in an unintended manner if the foot switch cover becomes damaged and impinges on the foot switch.

    • Terminated Z-2223-2012 FDA Class II initiated 18 Jul 2012

    1. The current system design of software version 4.0.0xx379 does not prevent the system from scanning if the dose level exceeds a set upper dose limit for CT Brain Perfusion scanning, which previous versions did. However, the current version does have the Dose Check (NEMA XR-25) feature, which provides dose notifications and alerts, but it is possible for the user to disable Dose Check after ente

    • Terminated Z-2035-2012 FDA Class II initiated 2 Jul 2012

    Philips Healthcare was notified that several of the sixteen bolts securing the brackets which support the carbon table top on various models of their Brilliant CT scanner have been found to be defective and the bolt heads may break off.

    • Terminated Z-1988-2012 FDA Class II initiated 23 May 2012

    Philips is issuing a software update to the Intellispace Portal systems to address a number of non-conformances. The most frequent and severe non-conformance identified involves the portal setting of "PAC Integration". This setting has a feature that allows the user to launch images of multiple patient studies into an application, but does not display a warning message informing the user that diff

    • Terminated Z-1987-2012 FDA Class II initiated 23 May 2012

    The Intellispace Portal software and Extended Brillance Workspace software may have significantly different results in the colors and size of the affected area when the customer uses the "Time Sensitive" and "Time Insensitive" methods.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Timeline

101 dated events

  1. Recall initiated FDA recalls Z-2174-2018
  2. Recall initiated FDA recalls Z-1143-2019
  3. Recall initiated FDA recalls Z-0543-2019
  4. Recall initiated FDA recalls Z-0544-2019
  5. Recall initiated FDA recalls Z-1982-2021
  6. Recall initiated FDA recalls Z-2603-2025
  7. Recall initiated FDA recalls Z-2604-2025
  8. Recall initiated FDA recalls Z-2595-2025
  9. Recall initiated FDA recalls Z-2602-2025
  10. Recall initiated FDA recalls Z-3043-2026

91 earlier events are in the sections above.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
JAK
Device class
Class II
Regulation
21 CFR Part 892.1750
Advisory committee
Radiology
FDA decision
SESE Substantially Equivalent

Similar devices

Other devices filed under FDA product code JAK (System, X-Ray, Tomography, Computed).