Pipeline Flex Embolization Device and Pipeline Flex Embolization Device with Shield Technology
by Micro Therapeutics, Inc. d/b/a ev3 Neurovascular
PMA P100018
Active
Valid until
No expiry
FDA recalls
4 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA PMAAPPR | NumberP100018 | Date6 Apr 2011 | StatusActive |
Recalls
-
- Ongoing Z-1245-2025
- Ongoing Z-1246-2025
Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically noted at 6-12 month imaging follow-up, potentially can lead to thrombosis and/or serious adverse events. Higher risk in females less-than-or-equal-to 45 years of age
-
- Ongoing Z-2403-2021
- Ongoing Z-2402-2021
Due to potential push wire fractures in the delivery system during use. The issue presents as a fracture in the spiral cut area of the Hypotube.
-
- Terminated Z-1527-2020
Embolization device delivery system may fracture at the distal section during device implantation. A weakened bond may result in unintended separation, where the distal portion of the device delivery system remains in the patient. If a Pipeline Flex embolization device has already been implanted successfully, there is no increased risk to patients due to this issue.
-
- Terminated Z-0309-2017
- Terminated Z-0308-2017
- Terminated Z-0307-2017
- Terminated Z-0310-2017
Delamination and detachment of the PTFE (polytetrafluoroethylene) coating material may lead to PTFE coating in the blood stream.
-
- Terminated Z-0120-2016
The firm is recalling Pipeline and Pipeline Flex Embolization Devices from U.S. since the devices were shipped with an EU version of the Instructions for Use.
-
- Terminated Z-1453-2014
Potential for the PTFE (polytetrafluoroethylene) coating to delaminate and detach from the delivery wire.
-
- Terminated Z-2045-2013
Covidien Neurovascular has initiated the recall of several lots of Pipeline Embolization Device (PED) because the Instructions For Use (IFU) do not include the following contraindication: Patients in whom a pre-existing stent is in place in the parent artery at the target aneurysm location. Additionally, the designation of aspirin and clopidogrel was missing from the following contraindication "
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA PMA Active
Specification
- Trade Name
- Pipeline Embolization Device
- Generic Name
- Intracranial aneurysm flow diverter
- PMA Number
- FDA Product Code
- OUT
- Device Class
- Class III
- Advisory Committee
- Neurology
- FDA Decision
- APPR Approved
Overview
AI-generated summary, 18 Sep 2026
The Pipeline Flex Embolization Device is an intracranial aneurysm flow diverter, classified as a Device Class 3 medical device. It is designed to promote aneurysm occlusion by diverting blood flow away from the aneurysm sac, thereby reducing pressure and facilitating natural healing processes in wide-necked intracranial aneurysms located in the internal carotid artery from the petrous to the superior hypophyseal segments.
This device is supplied as a single-use, endovascular implant and is intended for use in neurovascular procedures under radiographic guidance. It is approved for adults aged 22 and above and requires sterile handling, with specific sizing and deployment considerations based on anatomical requirements. The device is regulated under an FDA Premarket Approval (PMA) process, with the PMA number P100018, and is categorized under the product code OUT in the regulatory database.
Other devices from Micro Therapeutics, Inc. d/b/a ev3 Neurovascular
See all devices from Micro Therapeutics, Inc. d/b/a ev3 Neurovascular →
Questions
AI-generated answers, 18 Sep 2026
What is the Pipeline Flex Embolization Device used for in clinical practice?
The Pipeline Flex Embolization Device is designed to treat large or giant wide-necked intracranial aneurysms in adults aged 22 and above. It works by diverting blood flow away from the aneurysm sac in the internal carotid artery, reducing pressure and promoting natural healing. This is specifically for aneurysms located between the petrous and superior hypophyseal segments.
Is the Pipeline Flex Embolization Device intended for reuse, or is it single-use?
The Pipeline Flex Embolization Device is explicitly supplied as a single-use, endovascular implant. It requires sterile handling and cannot be reused after deployment, as is standard for most intracranial implants.
What regulatory pathway does the Pipeline Flex Embolization Device follow, and how can I verify its approval?
The device follows the FDA Premarket Approval (PMA) pathway, with the PMA number **P100018**. It was approved for its intended use in intracranial aneurysm treatment after undergoing a rigorous review process, including an expedited review. The approval notice was published in the Federal Register on **2011-04-20**.
Are there different sizes or models available for the Pipeline Flex Embolization Device?
Yes, the Pipeline Flex Embolization Device comes in multiple sizes, as indicated by the **FEI numbers** (e.g., 3010079067, 3007628272, etc.) listed in the regulatory records. These variations allow for customization based on anatomical requirements during deployment in the internal carotid artery.
How should the Pipeline Flex Embolization Device be stored before use?
Since it is a sterile, single-use device, storage conditions must maintain sterility until deployment. While specific storage instructions are not detailed here, typical best practices for endovascular implants—such as keeping it in a sealed, sterile packaging at room temperature away from direct sunlight—should be followed. Always refer to the manufacturer’s provided instructions for precise storage guidelines.