Pneupac Parapac Plus
by Smiths Medical International, Ltd.
510(k) K123957
Active
Valid until
No expiry
FDA recalls
6 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK123957 | Date16 Sep 2013 | StatusActive |
Recalls
-
- Ongoing Z-2973-2024
- Ongoing Z-2974-2024
There is a potential for inadvertent tidal volume knob movement from the original setting when set at high (1000-1500 mL) and low (70-150 mL) settings.
-
- Ongoing Z-3091-2024
- Ongoing Z-3090-2024
Smiths Medical became aware of an issue related to a potential for the patient outlet connector to loosen/detach from the paraPAC Plus P300 and P310 ventilators impacting the active ventilation function.
-
- Ongoing Z-1275-2024
- Ongoing Z-1274-2024
When a paraPAC plus ventilator is switched to the operating mode of VENTILATE, the ventilator may intermittently provide continuous positive gas flow instead of the intended cycling like a human breath. This non-cycling and continuous positive gas flow when in the cycling mode, is a malfunction, not allowing the ventilator to properly function as designed.
-
- Terminated Z-0163-2022
Two devices were labeled with the same serial number
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- BTL
- FDA applicant
- Smiths Medical International , Ltd.
- Device Class
- Class II
- Regulation
- 21 CFR Part 868.5925
- Advisory Committee
- Anesthesiology
- FDA Decision
- SESE Substantially Equivalent