Skip to content

Porous Patella

by Encore Medical, L.P. Dba Djo Surgical

510(k) K212941

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

1 ongoing

This device is named in the recall

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK212941 Date20 Jan 2022ClearedStatusActive

Recalls

1 recall · 1 ongoing · FDA

    • Ongoing Z-1384-2023 FDA Class II initiated 28 Feb 2023

    Patella drills are marked with incorrect diameter size of 5.6mm, when they should be 5.4mm. Porous patella overdrills are marked with incorrect diameter size of 5.4 mm, when they should be 5.6 mm. If the incorrect drill is selected during surgery, there is risk of the implant loosening, implant not properly fitting, and/or fractured bone.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
MBH
Device Class
Class II
Regulation
21 CFR Part 888.3565
Advisory Committee
Orthopedic
FDA Decision
SESE Substantially Equivalent