Skip to content

Port-A-Cath- II Dual-Lumen Low Profile System, Port-a-Cath II Dual Lumen Low Profile System with Introducer Set, Port-a

by Sims Deltec, Inc.

510(k) K974471

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

3 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK974471 Date22 Jan 1998ClearedStatusActive

Recalls

3 recalls · FDA

    • Terminated Z-1073-2018 FDA Class II initiated 4 Dec 2017

    Certain models and lots of PORT A CATH implantable port kits may have been misassembled. The kits may incorrectly contain a 6Fr or 6.5Fr size introducer instead of the correct 8.5Fr size introducer. The Power PORT A CATH II kit may also contain the incorrect locking collars in addition to the incorrect sized introducer.

  • 2 recalls, same reason

    • Terminated Z-0878-04 FDA Unclassified (pre-2012) initiated 29 Mar 2004
    • Terminated Z-0877-04 FDA Unclassified (pre-2012) initiated 29 Mar 2004

    Some of the outer product pouches have voids in the seals and do not assure that the outer surfaces of the inner tray containing the port, catheter and procedural components remain sterile after sterilization.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
LJT
Device Class
Class II
Regulation
21 CFR Part 880.5965
Advisory Committee
General Hospital
FDA Decision
SESE Substantially Equivalent

Other devices from Sims Deltec, Inc.