Portrait Core Services CSSXB
by GE Medical Systems Information Technologies, Inc. United States EUDAMEDGUDID
Ref 5973340
Product code BZQEMDN Z12030201
Regulatory status
Recall ongoing. 4 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | GUDID listing GUDID | Number00195278990914 | Date10 Dec 2025 | StatusIn distribution |
| European Union | CE/MDR EUDAMED | Number8406821BUG00424HE | Date17 Jan 2024 | StatusActive |
Coverage
- Active
- Listed
- Not covered
- European Union
- United States
Recalls
-
- Ongoing Z-3036-2026
- Ongoing Z-3037-2026
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.
-
- Ongoing Z-2878-2026
Portrait Core Services Software version V1.2.0 Security Vulnerability
-
- Ongoing Z-0918-2025
There is a software issue that can result in the loss of patient monitoring on the Portrait Central Viewer Application after 425 days of continuous run-time.
Specification
- Trade Name
- Portrait Core Services CSSXB EUDAMEDPortrait CSSXB GUDID
- Generic Name
- Bedside information system application software
- Device Name
- Portrait Core Services
- Model / Ref / Catalog
- 5973340, CSSXB
- Primary DI / UDI-DI
- 00195278990914
- Basic UDI-DI
- 8406821BUG00424HE
- 510(k) Number
- FDA Product Code
- BZQ, DQA, DRG, MSX, MWI
- EMDN Code
- Z12030201 Central information computing systems
- GMDN Term
- Bedside information system application software
- Device Class
- Class IIb EUDAMEDClass II GUDID
- Regulation Number
- 868.2375, 870.2700, 870.2910, 870.2300
- Market of Registration
- France
- Commercial Distribution
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Model / Reference | 5973340 | CSSXB |
| Device Class | Class IIb | Class II |
Manufacturer details
- EUDAMED SRN
- US-MF-000017529
- Authorised representative
- GE Medical Systems SCS FR-AR-000000344
Properties & raw data
Show properties, registry sources and raw records
Other properties (22)
- Sterile: No
- Single Use: No
- Implantable: No
- Reusable: No
- Contains Latex: No
- Reprocessed: No
- CMR Substance: No
- Human-derived: No
- Human Tissues: No
- Animal Tissues: No
- Endocrine Disruptor: No
- Administering Medicine: No
- Measuring Function: No
- Incorporates Medicinal Product: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Serial Number: No
- Expiration Date: No
- Manufacturing Date: No
Sources (2)
- EUDAMED 6ec5de5d-a64c-4cf4-8930-a251cf9ea561