Primus
by Siemens Medical Solutions USA, Inc.
510(k) K993425
Primus is a Class II device made by Siemens Medical Solutions USA, Inc., cleared under FDA 510(k) K993425 on 26 Nov 1999, active as of 22 Oct 2026. 50 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK993425 | Date26 Nov 1999 | StatusActive |
Official records
- FDA 510(k) database K993425
Recalls
-
- Terminated Z-1680-2020
Potential safety issue with the dose monitoring system safety interlocks which could result in a 10-20% overdose of radiation therapy.
-
- Terminated Z-1204-2019
Advisory issued not to disable or bypass the AMP function. Disabling or bypassing the AMP function for reasons of comfort is not advisable may lead to patient or user injury due to collision with LINAC equipment disabling the Auto Field Sequence (AFS) Automatic Motion Protection (AMP) function implemented at the control of the Digital Linear Accelerator.
-
- Terminated Z-1402-2018
- Terminated Z-1397-2018
Control Console software has been updated to reduce the risk for collision when delivering automatically sequenced treatments with automatic movements of the gantry and/or the treatment table.
-
- Terminated Z-1489-2017
Software update
-
- Terminated Z-1215-2017
Siemens became aware of Incorrect values for the rotational collimator position displayed due to mechanical problems of the sensor slide. In addition, there are improper weld seam at the overhead suspension.
-
- Terminated Z-1432-2017
Siemens became aware of a defective weld seam at overhead suspensions. Due to this defect it is possible for the weld seam to break and to cause the overhead suspension to fall down. This can lead to a severe injury to a patient or any other person.
-
- Terminated Z-2877-2016
Software patch installation to address several safety issues.
-
- Terminated Z-2812-2015
A software fix has been released to prevent automatic movement resulting in a collision safety risk for patients.
-
- Terminated Z-0168-2016
potential safety risk of collision of the LINAC gantry with the patient or the treatment table in case of automatically sequenced treatment delivery techniques using SIMTEC" Auto Field Sequence Option.
-
- Terminated Z-1059-2015
A hardware update to replace the old PCB Assembly, Head Driver-G41 with PCB with overcurrent protection as a result of reports of the electric board inside the gantry showing signs of overheating, including thermal damage and smoke emission.
-
- Terminated Z-0980-2015
Use of any unauthorized third-party components on Siemens LINACs may lead to patient or operator injury, damage to the LINAC, and can potentially compromise customer warranty. The dark current dose must be considered in the overall dose calculation, or an overdose and serious injury may occur.
-
- Terminated Z-2095-2013
Siemens has become aware of a customer complaint concerning the rail threaded holes of the 550TxT treatment tabletop. Due to a manufacturing error, the screws holding the side rails in position on the tabletop might break off.
-
- Terminated Z-1663-2013
The information regarding the FILM Mode calibration may have been considered difficult to interpret when translated. SIEMENS HEALTHCARE CR / RADIATION ONCOLOGY had become aware that this issue may have resulted in mistreatment of patients.
-
- Terminated Z-1182-2013
Software update to fix multiple safety related issues.
-
- Terminated Z-1494-2012
The firm became aware of a potential issue with Siemens-branded ONCOR or PRIMUS linear accelerator with an OPTIVUE or MVision imaging system that has the potential collision problem of the imaging panel if the panel does not retract before the gantry rotates; this may result in patient injury.
-
- Terminated Z-1092-2012
Software upgrade to installed base to the latest release of the RT Therapist version 4.1.144 which corrects multiple safety issues that could lead to patient mistreatment.
-
- Terminated Z-0782-2012
Bolts holding the trim weight may break causing the plates of the counterweight to become loose and fall into the cover.
-
- Terminated Z-3003-2011
A potential safety issue was discovered when a pause for an imaging segment is inserted (where an imaging segment does not exist) with the ARTISTE and RT Therapist 4.1 system will continue to the next image.
-
- Terminated Z-2789-2011
Potential safety issue of possible data loss for the transfer of the same patients treatment data between the LANTIS OIS and the RTT v2.x software and the RT Therapist version 4.1 system
-
- Terminated Z-2243-2011
Siemens found a potential safety issue related to patient treatment with table offsets on syngo RT Therapist (RTT) 4.1 and 4.2 which could result in dose being delivered to wrong location.
-
- Terminated Z-2216-2011
Siemens has become aware of a potential safety issue that may result in mistreatment when customers are configured with the ARTISTE Linac RT Therapist v4.2 in combination with RT Oncologist v.4.2 and use the remote "Online 2D Image Review.
-
- Terminated Z-2190-2011
When using the ARTISTE Linac and RT Therapist v4.1 along with an imported cropped reference image, the image maybe processed incorrectly, thus resulting in a mistreatment.
-
- Terminated Z-2640-2011
- Terminated Z-2643-2011
- Terminated Z-2644-2011
- Terminated Z-2641-2011
The firm has become aware of a potential issue that may result in a service engineer receiving an electrical shock should the connector cable housing become unfastened from the chassis.
-
- Terminated Z-1970-2011
When editing the note type of a previously entered note, then reverting to the original type, data may be lost.
-
- Terminated Z-1396-2008
- Terminated Z-1398-2008
- Terminated Z-1397-2008
Linear accelerator gantry may rotate in an unexpected direction and may cause patient injury on impact.
-
- Terminated Z-1759-2008
Radiation leak: When used with field sizes of 5cm x 5cm or smaller, product may leak radiation at a distance of 2 cm from the side of the applicator body up to 13%, in excess of IEC standards.
-
- Terminated Z-0686-2008
- Terminated Z-0687-2008
Incorrect field size -- Products manufactured prior to 1999 may experience jaw field size errors during interlock resulting in potential mistreatment if a light field check is not performed. This problem is present in machines with serial numbers less than 3094, as they do not contain redundant sensors for jaw position.
-
- Terminated Z-0850-2008
Door Hinge failures; loose or falling off (Stationary Structure Doors)
-
- Terminated Z-0380-2008
Product's collision detection system may not be sensitive enough to trigger a stop action without a significant amount of force.
-
- Terminated Z-0166-2008
- Terminated Z-0167-2008
- Terminated Z-0169-2008
- Terminated Z-0168-2008
Mistreatment of radiation therapy due to inadvertent parameter changes or miss-calibration in Service Mode. The miss-calibration can affect the beam profile and Virtual Wedge angle. This leads to patient mistreatment and alteration of future treatment plans. A second reason is an unplanned Dosimetry calibration data set can be erroneously copied during "Service Mode."
-
- Terminated Z-0355-06
Incomplete installation of the optional Beam Shield may allow unexpected doses of radiation outside the radiation protected areas.
-
- Terminated Z-1044-2013
- Terminated Z-1043-2013
- Terminated Z-1041-2013
- Terminated Z-1036-2013
- Terminated Z-1040-2013
- Terminated Z-1037-2013
- Terminated Z-1039-2013
- Terminated Z-1038-2013
- Terminated Z-1042-2013
Siemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance defects and several safety issues that were not considered critical (for immediate correction) for the safe use of the device.
Timeline
- Recall initiated FDA recalls Z-0168-2016
- Recall initiated FDA recalls Z-2812-2015
- Recall initiated FDA recalls Z-2877-2016
- Recall initiated FDA recalls Z-1215-2017
- Recall initiated FDA recalls Z-1432-2017
- Recall initiated FDA recalls Z-1489-2017
- Recall initiated FDA recalls Z-1402-2018
- Recall initiated FDA recalls Z-1397-2018
- Recall initiated FDA recalls Z-1204-2019
- Recall initiated FDA recalls Z-1680-2020
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- IYE
- Device class
- Class II
- Regulation
- 21 CFR Part 892.5050
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code IYE (Accelerator, Linear, Medical).
- Dose Calculator by Sun Nuclear Corporation
- KLARITY MEDICAL R7302-15C by Klarity Medical Products, LLC
- XKNIFE-3 by Radionics Software Applications, Inc.
- AKTINA MEDICAL PHYSICS CORPORATION TILT BOARD by Aktina Medical Physics Corp.
- MILLCOMP by MATH RESOLUTIONS, LLC
- CLINAC (R) C SERIES (400C, 600C, 2100C, 2500C) by Varian Assoc., Inc.
- TrueBeam, TrueBeam STx, Edge by Varian Medical Systems, Inc.
- KLARITY MEDICAL RNB-10 by Klarity Medical Products, LLC
- BEAM BLOCK TRAYS by Arplay Medical S.A.
- Halcyon Ethos Radiotherapy System (5.0) by Varian Medical Systems, Inc.