Skip to content

Prismaflex Set 107640US

by Gambro Industries

Prismaflex Set 107640US is a haemofilter (FDA Class II; FDA product code KDI, Dialyzer, high permeability with or without sealed dialysate system) made by Gambro Industries, cleared under FDA 510(k) K212216 on 1 Apr 2022 and listed in FDA GUDID since 22 Nov 2022 (UDI-DI 00085412917696), active as of 23 Oct 2026. 3 FDA recalls are linked to it.

Active United States. Also listed in GUDID (US).

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

3 ongoing

This device is named in the recall

Documents

510(k) K212216

FDA submission, from GUDID

Trust

Unverified

Source GUDID, synced 23 Oct 2026.

Regulatory status

as of registry sync 23 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number00085412917696 UDI-DI Date22 Nov 2022ListedStatusListed GUDID: in commercial distribution
United StatesFDA 510(k)SESE from GUDID Substantially Equivalent · Traditional NumberK212216 Date1 Apr 2022DecidedStatusActive

Official records

From the FDA record

Device description
Disposable extracorporeal circuit
MRI safety
Labeling does not contain MRI Safety Information
Natural rubber latex
Labelled as not made with natural rubber latex
Sterility
Supplied sterile

From the GUDID record (AccessGUDID), as read on 23 Oct 2026, as the labeller entered it.

Across registries

2 registries list this device

  • FDA 510(k) K212216 decided 1 Apr 2022 Active
  • FDA GUDID UDI-DI 00085412917696 listed 22 Nov 2022 Listed

Recalls

3 recalls · 3 ongoing · FDA

  • 2 recalls, same reason

    • Ongoing Z-1224-2026 FDA Class II initiated 6 Jan 2026
    • Ongoing Z-1225-2026 FDA Class II initiated 6 Jan 2026

    There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.

    • Ongoing Z-0084-2023 FDA Class II initiated 14 Sep 2022

    Within the current Prismaflex set IFU (instructions for use), there is a mistranslation of the Extonian (Eesti) text. The mistranslation indicates contradictory information related to the patient body weight restrictions.

Timeline

5 dated events

  1. Cleared FDA 510(k) K212216
  2. Recall initiated FDA recalls Z-0084-2023
  3. Listed FDA GUDID 00085412917696
  4. Recall initiated FDA recalls Z-1224-2026
  5. Recall initiated FDA recalls Z-1225-2026

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

GUDID

Trade name
Prismaflex Set
Generic name
Haemofilter
Model / Ref / Catalog
107640US
Primary DI / UDI-DI
00085412917696
510(k) Number
FDA product code
KDI
GMDN term
Haemofilter
Device class
Class II
Regulation number
876.5860
GUDID distribution status
In Commercial Distribution

Similar devices

Other devices filed under FDA product code KDI (Dialyzer, high permeability with or without sealed dialysate system).

Other devices from Gambro Industries

  • Prismaflex Set 107636US Open recall

    Class II
  • Prismaflex Set HF1000 Open recall

    Class II
  • Prismaflex Set M150 Open recall

    Class II
  • Diaclear 106887

    Class II

See all devices from Gambro Industries →

Properties

  • Supplied sterile
  • Single use
  • Prescription use
  • Lot / batch number
  • Expiration date
  • Manufacturing date
Other properties (9, all answered No)