PrisMax System Version 3
by Baxter Healthcare Corporation 510(k)
Active
510(k) K193482
Regulatory status
Recall ongoing. 2 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK193482 | Date13 Mar 2020 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-2438-2025
Unexpected PrisMax System alarm T0830: Blood Leak Detected, when no blood leak is present, or difficulty normalizing the Blood Leak Detector (BLD) leading to PrisMax System alarms T1313: BLD Normalize Failed, T0853: Normalization Failed, or T1205: BLD Self-Test Failure
-
- Ongoing Z-2439-2025
Thermax System may be unable to detect the presence of a bag on the Thermax Blood Warmer Unit leading to difficulty setting up therapy, or PrisMax System alarm T2284: Thermax Disposable Not Inserted
Specification
- 510(k) Number
- FDA Product Code
- KDI
- Device Class
- Class II
- Regulation
- 21 CFR Part 876.5860
- Advisory Committee
- Gastroenterology, Urology
- FDA Decision
- SESE Substantially Equivalent