Skip to content

Probeat-FR

by Hitachi High-Tech Corporation

Probeat-FR is a proton therapy system (FDA Class II; FDA product code LHN, System, Radiation Therapy, Charged-Particle, Medical) made by Hitachi High-Tech Corporation, cleared under FDA 510(k) K232032 on 12 Jan 2024 and listed in FDA GUDID since 1 Apr 2026 (UDI-DI 04560333350440), active as of 23 Oct 2026. 4 FDA recalls are linked to it.

Active United States. Also listed in GUDID (US).

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

4 ongoing

This device is named in the recall

Documents

510(k) K232032

FDA submission, from GUDID

Trust

Unverified

Source GUDID, synced 23 Oct 2026.

Regulatory status

as of registry sync 23 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number04560333350440 UDI-DI Date1 Apr 2026ListedStatusListed GUDID: in commercial distribution
United StatesFDA 510(k)SESE from GUDID Substantially Equivalent · Traditional NumberK232032 Date12 Jan 2024DecidedStatusActive

Official records

From the FDA record

Device description
The PROBEAT-FR is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose, dose distribution and directed to the prescribed patient treatment site. The PROBEAT-FR is a modification to the cleared PROBEAT-CR system, for installation at a different clinical site.
MRI safety
Labeling does not contain MRI Safety Information
Natural rubber latex
No latex statement on the label
Sterility
Supplied non-sterile

From the GUDID record (AccessGUDID), as read on 23 Oct 2026, as the labeller entered it.

Across registries

2 registries list this device

  • FDA 510(k) K232032 decided 12 Jan 2024 Active
  • FDA GUDID UDI-DI 04560333350440 listed 1 Apr 2026 Listed

Recalls

4 recalls · 4 ongoing · FDA

    • Ongoing Z-1429-2026 FDA Class II initiated 6 Feb 2026

    Software anomaly in the patient positioning system may result in positional discrepancy.

  • 3 recalls, same reason

    • Ongoing Z-3126-2024 FDA Class II initiated 27 May 2024
    • Ongoing Z-3125-2024 FDA Class II initiated 27 May 2024
    • Ongoing Z-3127-2024 FDA Class II initiated 27 May 2024

    During positioning within the system's 3D to 3D matching mode while transferring CBCT images, when the Float Image Selection button is pressed, it may display CBCT images that are insufficient in number.

Timeline

6 dated events

  1. Cleared FDA 510(k) K232032
  2. Recall initiated FDA recalls Z-3126-2024
  3. Recall initiated FDA recalls Z-3125-2024
  4. Recall initiated FDA recalls Z-3127-2024
  5. Recall initiated FDA recalls Z-1429-2026
  6. Listed FDA GUDID 04560333350440

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

GUDID

Trade name
Probeat
Generic name
Proton therapy system
Model / reference
PROBEAT-FR
Primary DI / UDI-DI
04560333350440
510(k) Number
FDA product code
LHN
GMDN term
Proton therapy system
Device class
Class II
Regulation number
892.5050
GUDID distribution status
In Commercial Distribution

Similar devices

Other devices filed under FDA product code LHN (System, Radiation Therapy, Charged-Particle, Medical).

Other devices from Hitachi High-Tech Corporation

  • Probeat-Cr

    Class II
  • Probeat

    Class II
  • Probeat with Mgcs

    Class II
  • Probeat-V

    Class II

See all devices from Hitachi High-Tech Corporation →

Properties

  • Prescription use
  • Serial number
  • Manufacturing date
Other properties (12, all answered No)