ProBP 2400 Digital Blood Pressure Device
by Microlife Intellectual Property GmbH 510(k)
Active
510(k) K152770
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK152770 | Date21 Oct 2015 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-1777-2017
A defective electrical component in the battery charging circuit in two lots of the ProBP 2400 may potentially cause an over-voltage battery charging condition when the device is connected to the external power supply/battery charger. This potential defect resides with the device, not the battery. The over-voltage condition can result in high NiMH battery temperatures that, in some cases, can reach levels sufficient to cause melting of the plastic battery door and other adjacent plastic and foam device components.
Specification
- 510(k) Number
- FDA Product Code
- DXN
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.1130
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent