Prometra Programmable Infusion Pump System
PMA P080012
Active
Valid until
No expiry
FDA recalls
6 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA PMAAPPR | NumberP080012 | Date7 Feb 2012 | StatusActive |
Recalls
-
- Ongoing Z-1124-2020
- Ongoing Z-1122-2020
- Ongoing Z-1123-2020
- Ongoing Z-1120-2020
- Ongoing Z-1125-2020
- Ongoing Z-1121-2020
A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system.
-
- Terminated Z-1996-2021
Physician Order Form (PL-15400-00 Nov 2018) distributed is discontinued due to the text 'Refill the pump and program refill with the same medication removed prior to MRI' may be misinterpreted as 'reuse the same drug solution extracted from the pump prior to the MRI procedure', creating a possible risk of infection
-
- Terminated Z-1714-2019
- Terminated Z-1715-2019
- Terminated Z-1716-2019
The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician Programmer, Software Version 2.00.29, to program a Bridge or Demand bolus while the pump is set to either a Periodic Flow or Multiple Rates flow mode. Error 115 can also occur if dendrites form or if the pump is dropped.
-
- Terminated Z-1712-2019
- Terminated Z-1713-2019
The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician Programmer, Software Version 2.00.29, to program a Bridge or Demand bolus while the pump is set to either a Periodic Flow or Multiple Rates flow mode.
-
- Terminated Z-2740-2017
Flowonix Medical received a report of a patient implanted with the Prometra II Programmable Pump who may have received a fatal drug overdose during an MRI procedure. The Prometra II pump has an FDA-approved design feature intended to permit safe exposure to an MRI without removing drug from the reservoir.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA PMA Active
Specification
- Generic Name
- Pump, infusion, implanted, programmable
- PMA Number
- FDA Product Code
- LKK
- Device Class
- Class III
- Advisory Committee
- General Hospital
- FDA Decision
- APPR Approved
Overview
AI-generated summary, 18 Sep 2026
The Prometra Programmable Infusion Pump System is an implanted, programmable infusion pump. It is indicated for intrathecal infusion of Infumorph (preservative-free morphine sulfate sterile solution) or preservative-free sterile 0.9% saline solution.
The device is supplied as an implanted system and is intended for use in general hospital settings. It has received FDA PMA approval (P080012) and is classified as a Class III device. Sterility, MRI safety, storage conditions, and reusability are not specified in the provided data.
Questions
AI-generated answers, 18 Sep 2026
What is the Prometra Programmable Infusion Pump System indicated for?
The device is indicated for intrathecal infusion of Infumorph (preservative-free morphine sulfate sterile solution) or preservative-free sterile 0.9% saline solution, as stated in the approval documentation.
Who manufactures the Prometra Programmable Infusion Pump System?
The Prometra Programmable Infusion Pump System is manufactured by Infusyn Therapeutics, LLC, located in Hackettstown, New Jersey.
What regulatory authorization does the Prometra Programmable Infusion Pump System have?
The device has received FDA Premarket Approval (PMA) under number P080012, with approval granted on February 7, 2012.