United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK132084 Date26 Sep 2013ClearedStatusActive

Recalls

2 recalls · FDA

    • Terminated Z-0671-2018 FDA Class II initiated 8 Nov 2017

    When an error message remains displayed and is not cleared in the Protura software with the Elekta pedestal coordinates, moving the pedestal could cause the Protura software to not update with the pedestal location and buffers the pedestal movement history.

    • Terminated Z-1992-2016 FDA Class II initiated 31 Mar 2016

    When using the recalled product in split shift mode, the non-sub-millimeter positioning has been found to be inconsistent with the setting parameters.