Pvs 2000 SYNCHRO2 Guidewire
by Boston Scientific - Precision Vascular 510(k)
Active
510(k) K053268
Regulatory status
Recall ongoing. 1 ongoing FDA recall. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK053268 | Date13 Mar 2006 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-1941-2024
Stryker Neurovascular has observed an increased frequency of PTFE coating damage occurring on the Synchro Guidewires that may be caused by the practice of backloading the guidewire through the optional introducer accessory. This issue is limited to certain lots of the Synchro Guidewire that contain an older version of the introducer accessory. Due to variation in the manufacturing process of the supplier of the introducer accessory, certain lots of introducers have sharper than intended edges that can peel off the PTFE coating when physicians use a technique known as backloading. Users with impacted product in their inventory are cautioned not to use this backloading technique.
-
- Terminated Z-0868-2011
- Terminated Z-0870-2011
- Terminated Z-0873-2011
- Terminated Z-0871-2011
- Terminated Z-0875-2011
- Terminated Z-0869-2011
- Terminated Z-0872-2011
- Terminated Z-0874-2011
The Guidewire has the potential to flake off polytetrafluroethylene (PTFE) particulate, which has the potential to embolize and result in infarction when introduced into a patient's circulatory system.
-
- Terminated Z-2070-2010
- Terminated Z-2071-2010
- Terminated Z-2068-2010
- Terminated Z-2069-2010
Boston Scientific is initiating a Medical Device Recall regarding specific lots of the FathomTM-14 Steerable Guidewire. For lubricity, the distal portion of these guidewires is coated with hydrophilic polymer and the proximal portion is coated with polytetrafluoroethylene (PTFE). Through internal inspection process, the firm determined that Fathom guidewires from affected lots may exhibit a P
Specification
- 510(k) Number
- FDA Product Code
- DQX
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.1330
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent