Recalls
-
- Terminated Z-1255-2021
R&D Systems, Inc. received two complaints that the standard signal was low and controls were not within the specification range provided with the kit. Investigation confirmed low standard signal and controls running out of specification for kits returned from the customer and the retained kits.
-
- Terminated Z-0691-2019
The Human sTfR Quantikine IVD ELISA Kit, catalog # DTFR1, lot P180151 was packed with the incorrect control data card number 750228.4.
-
- Terminated Z-2887-2016
R&D Systems, Inc. received two customer complaints that results of the Controls provided with the Quantikine IVD Human sTfR Immunoassay kit were out of the range high. Internal testing within R&D Systems confirmed a problem with Control results being out of range high.