Skip to content

Quantum Blast

by Pride Mobility Products Corporation

510(k) K011993

Quantum Blast is a medical device (FDA Class II; FDA product code ITI, Wheelchair, Powered) made by Pride Mobility Products Corporation, cleared under FDA 510(k) K011993 on 13 Jul 2001, active as of 8 Nov 2026. 3 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

2 ongoing

This device is named in the recall

Documents

Trust

Unverified

Source FDA 510(k), synced 8 Nov 2026.

Regulatory status

as of registry sync 8 Nov 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Special NumberK011993 Date13 Jul 2001ClearedStatusActive

Official records

From the FDA record

Received by the FDA
26 Jun 2001
FDA decision
13 Jul 2001 (Substantially Equivalent)
FDA review time
17 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K011993 (openFDA), as read on 8 Nov 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

3 recalls · 2 ongoing · FDA

  • 2 recalls, same reason

    • Ongoing Z-2844-2026 FDA Class II initiated 25 Jun 2026
    • Ongoing Z-2853-2026 FDA Class II initiated 25 Jun 2026

    If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.

    • Terminated Z-1776-2008 FDA Unclassified (pre-2012) initiated 11 Apr 2008

    Software - Watch dog timer feature was disabled.

Timeline

4 dated events

  1. Cleared FDA 510(k) K011993
  2. Recall initiated FDA recalls Z-1776-2008
  3. Recall initiated FDA recalls Z-2844-2026
  4. Recall initiated FDA recalls Z-2853-2026

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
ITI
FDA applicant
Pride Mobility Products Corp.
Device class
Class II
Regulation
21 CFR Part 890.3860
Advisory committee
Physical Medicine
FDA decision
SESE Substantially Equivalent

Similar devices

Other devices filed under FDA product code ITI (Wheelchair, Powered).

Other devices from Pride Mobility Products Corporation