Skip to content

RayCare 4.0

by RaySearch Laboratories AB (PUBL)

RayCare 4.0 is a radiology information system application software (FDA Class II; FDA product code MUJ, System, Planning, Radiation Therapy Treatment) made by RaySearch Laboratories AB (PUBL), cleared under FDA 510(k) K200487 on 10 Jun 2020 and listed in FDA GUDID since 15 Sep 2020 (UDI-DI 07350002010341), active as of 23 Oct 2026. 5 FDA recalls are linked to it.

Active United States. Also listed in GUDID (US).

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

4 ongoing

This device is named in the recall

Documents

510(k) K200487

FDA submission, from GUDID

Trust

Unverified

Source GUDID, synced 23 Oct 2026.

Regulatory status

as of registry sync 23 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesGUDID listing Number07350002010341 UDI-DI Date15 Sep 2020ListedStatusListed GUDID: in commercial distribution
United StatesFDA 510(k)SESE from GUDID Substantially Equivalent · Traditional NumberK200487 Date10 Jun 2020DecidedStatusActive

Official records

From the FDA record

MRI safety
Labeling does not contain MRI Safety Information
Natural rubber latex
No latex statement on the label
Sterility
Supplied non-sterile

From the GUDID record (AccessGUDID), as read on 23 Oct 2026, as the labeller entered it.

Across registries

2 registries list this device

  • FDA 510(k) K200487 decided 10 Jun 2020 Active
  • FDA GUDID UDI-DI 07350002010341 listed 15 Sep 2020 Listed

Recalls

5 recalls · 4 ongoing · FDA

    • Ongoing Z-0206-2024 FDA Class II initiated 12 Sep 2023

    An issue was found in RayCare 5B, 6A, including service packs, where values will not be visible for documents in a read-only state when opened outside the Documents workspace when using document fields configured with single- or multi-selectable values.

    • Ongoing Z-2464-2023 FDA Class II initiated 17 Jul 2023

    An issue was found in Software RayCare 5A, 5B, 6A, including service packs, where an allergy warning against medication substance (drug ingredient) will not be displayed as expected under certain circumstances

    • Ongoing Z-1281-2021 FDA Class II initiated 3 Mar 2021

    Patient related messages created in RayCare 3B, RayCare 4A will be lost when upgrading from RayCare 2 to RayCare 3B or later

    • Ongoing Z-2899-2020 FDA Class II initiated 29 Jul 2020

    When performing offline image review in RayCare 2C, RayCare 3A and RayCare 3B an offline image review task will guide you to the image viewer with the treatment image related to the fraction and the planning image selected. When starting the task, an error message might be displayed stating: Review files not available.

    • Terminated Z-2138-2021 FDA Class II initiated 8 Jun 2021

    Initial delivery positions will be set incorrectly when having setup beam(s) at the localization point.

Timeline

7 dated events

  1. Cleared FDA 510(k) K200487
  2. Recall initiated FDA recalls Z-2899-2020
  3. Listed FDA GUDID 07350002010341
  4. Recall initiated FDA recalls Z-1281-2021
  5. Recall initiated FDA recalls Z-2138-2021
  6. Recall initiated FDA recalls Z-2464-2023
  7. Recall initiated FDA recalls Z-0206-2024

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

GUDID

Trade name
RayCare
Generic name
Radiology information system application software
Model / reference
4.0
Primary DI / UDI-DI
07350002010341
510(k) Number
FDA product code
MUJ
GMDN term
Radiology information system application software
Device class
Class II
Regulation number
892.5050
GUDID distribution status
In Commercial Distribution

Similar devices

Other devices filed under FDA product code MUJ (System, Planning, Radiation Therapy Treatment).

Other devices from RaySearch Laboratories AB (PUBL)

See all devices from RaySearch Laboratories AB (PUBL) →

Properties

  • Prescription use
  • Manufacturing date
Other properties (13, all answered No)