  "format": "devicerts.device.v1",
  "cite_as": "RayStation 10.1, RaySearch Laboratories AB (PUBL). deviCERTS medical device catalogue. https://devicerts.com/devices/raystation-10-1-muj (data as of 2026-11-20).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/raystation-10-1-muj",
    "json_url": "https://devicerts.com/devices/raystation-10-1-muj.json",
    "markdown_url": "https://devicerts.com/devices/raystation-10-1-muj.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-20",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/raystation-10-1-muj",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "RayStation 10.1",
    "registry_name": "RayStation 10.1",
    "url": "https://devicerts.com/devices/raystation-10-1-muj",
    "manufacturer": {
      "name": "RaySearch Laboratories AB (PUBL)",
      "url": "https://devicerts.com/manufacturers/raysearch-laboratories-ab-publ",
      "country": "Sweden",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K210645",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "MUJ"
    ],
    "emdn": [],
    "gmdn": [],
    "description": {
      "text": "RayStation 10.1 is a Radiation Therapy Treatment Planning System designed for advanced radiotherapy planning. It integrates multi-criteria optimization and adaptive therapy capabilities to support external beam therapy, brachytherapy, and particle therapy (including protons, helium, and carbon ions). The system facilitates precise dose calculations and treatment planning across diverse radiation modalities, enhancing treatment accuracy and customization for complex cases.\n\nThis software-based system is intended for use in radiology departments and oncology centers, where it aids clinicians in generating optimized treatment plans. Supplied as a standalone software solution, it requires compatible hardware and network infrastructure for operation. The device holds FDA 510(k) clearance (K210645) under the Radiology advisory committee, classified as a Class II medical device under regulation 892.5050. No sterility or single-use requirements apply, as it is a reusable digital platform.",
      "ai_generated": true,
      "generated_on": "2026-10-01",
      "note": "Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data is the record."
    },
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K210645",
      "number_label": null,
      "device_udi_di": null,
      "date": "2021-06-29",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-1102-2026",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-11-28",
      "reason": "Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.",
      "applies": "this device",
      "via": "K210645"
    },
    {
      "number": "Z-2490-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-08-08",
      "reason": "DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.",
      "applies": "this device",
      "via": "K210645"
    },
    {
      "number": "Z-1188-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-01-16",
      "reason": "Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.",
      "applies": "this device",
      "via": "K210645"
    },
    {
      "number": "Z-1747-2024",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-03-28",
      "reason": "Potential for reported SSD to be too high.",
      "applies": "this device",
      "via": "K210645"
    },
    {
      "number": "Z-1482-2024",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-02-23",
      "reason": "A use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant.",
      "applies": "this device",
      "via": "K210645"
    },
    {
      "number": "Z-0848-2022",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2021-09-09",
      "reason": "If the dynamic jaw mode is used and an optimization is continued after changing the dose grid or modifying the target ROI, the positions of the jaws may change unexpectedly for some control points. User must be aware  to avoid unnecessary jaw openings during\ntreatment planning",
      "applies": "this device",
      "via": "K210645"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K210645",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "MUJ",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 892.5050",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Radiology",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}