# RayStation 11B SPT1

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/raystation-11b-spt1>
- Manufacturer: [RaySearch Laboratories AB (PUBL)](https://devicerts.com/manufacturers/raysearch-laboratories-ab-publ) (Sweden)
- Device class: Class II (US FDA)
- FDA product code: MUJ
- GMDN: Radiation therapy software (40887)

## Registrations and authorizations

- **GUDID** · GUDID listing · United States · UDI-DI 07350002010570 · listed 2023-05-12 · status: Listed (GUDID: in commercial distribution)
- **GUDID** · FDA 510(k) · United States · K220141 · decision 2022-04-15 · decision SESE · status: Active · from GUDID

## Recalls

- Z-1104-2026 · Class II · ongoing · initiated 2025-11-28: Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
- Z-2492-2025 · Class II · ongoing · initiated 2025-08-08: DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
- Z-1190-2025 · Class II · ongoing · initiated 2025-01-16: Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.
- Z-1749-2024 · Class II · ongoing · initiated 2024-03-28: Potential for reported SSD to be too high.
- Z-1482-2024 · Class II · ongoing · initiated 2024-02-23: A use error that has occurred with RayStation/RayPlan, where material override using Silicon, Si, was incorrectly selected for a silicone gel polymer implant.
- Z-1299-2022 · Class II · terminated · initiated 2022-06-10: This notice concerns two issues found related to the display of Linear Energy Transfer (LET) in RayStation 11B including some service packs. First, when using a dose threshold for an evaluation dose, LET display might be misleading. Second, a displayed beam-specific LET distribution can sometimes be out of sync with the selected beam.

## Identity

- Trade name: RayStation [FDA GUDID (AccessGUDID)]
- Generic name: Radiation therapy software [FDA GUDID (AccessGUDID)]
- Model / reference: 11B SPT1 [FDA GUDID (AccessGUDID)]

## Identifiers

- Primary DI / UDI-DI: 07350002010570 [FDA GUDID (AccessGUDID)]
- 510(k) Number: K220141 [FDA GUDID (AccessGUDID)]
- FDA product code: MUJ [FDA GUDID (AccessGUDID)]
- GMDN term: Radiation therapy software (40887) [FDA GUDID (AccessGUDID)]

## Classification

- Device class: Class II [FDA GUDID (AccessGUDID)]
- Regulation number: 892.5050 [FDA GUDID (AccessGUDID)]
- GUDID distribution status: In Commercial Distribution [FDA GUDID (AccessGUDID)]

## Properties

- Prescription use: Yes [FDA GUDID (AccessGUDID)]
- Manufacturing date: Yes [FDA GUDID (AccessGUDID)]
- Supplied sterile: No [FDA GUDID (AccessGUDID)]
- Sterilise before use: No [FDA GUDID (AccessGUDID)]
- Single use: No [FDA GUDID (AccessGUDID)]
- Implantable: No [FDA GUDID (AccessGUDID)]
- Contains latex: No [FDA GUDID (AccessGUDID)]
- Over-the-counter: No [FDA GUDID (AccessGUDID)]
- Convenience kit: No [FDA GUDID (AccessGUDID)]
- Combination product: No [FDA GUDID (AccessGUDID)]
- HCT/P: No [FDA GUDID (AccessGUDID)]
- Premarket exempt: No [FDA GUDID (AccessGUDID)]
- Lot / batch number: No [FDA GUDID (AccessGUDID)]
- Serial number: No [FDA GUDID (AccessGUDID)]
- Expiration date: No [FDA GUDID (AccessGUDID)]

## Source and citation

- Cite as: RayStation 11B SPT1, RaySearch Laboratories AB (PUBL). deviCERTS medical device catalogue. https://devicerts.com/devices/raystation-11b-spt1 (data as of 2026-11-20).
- Data sources: [FDA GUDID (AccessGUDID)](https://accessgudid.nlm.nih.gov/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-23
- Statuses computed as of: 2026-11-20
- JSON: <https://devicerts.com/devices/raystation-11b-spt1.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/raystation-11b-spt1
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.