  "format": "devicerts.device.v1",
  "cite_as": "RayStation 7, RaySearch Laboratories AB (PUBL). deviCERTS medical device catalogue. https://devicerts.com/devices/raystation-7-muj (data as of 2026-11-20).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/raystation-7-muj",
    "json_url": "https://devicerts.com/devices/raystation-7-muj.json",
    "markdown_url": "https://devicerts.com/devices/raystation-7-muj.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-20",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/raystation-7-muj",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "RayStation 7",
    "registry_name": "RayStation 7",
    "url": "https://devicerts.com/devices/raystation-7-muj",
    "manufacturer": {
      "name": "RaySearch Laboratories AB (PUBL)",
      "url": "https://devicerts.com/manufacturers/raysearch-laboratories-ab-publ",
      "country": "Sweden",
      "country_basis": null,
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K180379",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "MUJ"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K180379",
      "number_label": null,
      "device_udi_di": null,
      "date": "2018-04-25",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-2487-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-08-08",
      "reason": "DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.",
      "applies": "this device",
      "via": "K180379"
    },
    {
      "number": "Z-1185-2025",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2025-01-16",
      "reason": "Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.",
      "applies": "this device",
      "via": "K180379"
    },
    {
      "number": "Z-1742-2024",
      "classification": "Class II",
      "status": "ongoing",
      "initiated": "2024-03-28",
      "reason": "Potential for reported SSD to be too high.",
      "applies": "this device",
      "via": "K180379"
    },
    {
      "number": "Z-0679-2019",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-11-12",
      "reason": "Isocenter shifts when using dose tracking in RayStation 3.5,\nRayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayStation 7 and RayStation\n8A. In these versions, isocenter shifts in dose tracking are incorrectly interpreted according to the DICOM\npatient coordinate system.",
      "applies": "this device",
      "via": "K180379"
    },
    {
      "number": "Z-0294-2019",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-09-21",
      "reason": "Robust optimization for plans using beam set + background dose is not supported in RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6 and RayStation 7.",
      "applies": "this device",
      "via": "K180379"
    },
    {
      "number": "Z-2497-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-06-29",
      "reason": "If the beam model has a highly asymmetric primary source, it is not correctly taken into account in the calculation of DMLC fields when\nthe collimator is rotated. This could lead to potentially significant overdosage at delivery. The user must be aware of the issue to avoid incorrect dose calculations during treatment planning.",
      "applies": "this device",
      "via": "K180379"
    },
    {
      "number": "Z-2554-2018",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2018-06-06",
      "reason": "The firm has learned that some RayStation/RayPlan users have commissioned machines with erroneous Beam profile correction parameters. These parameters\naffect the dose calculated in corners of large or off-axis fields. This effect cannot be seen in the Beam Commissioning module and dose in large or off-axis fields needs to be validated using the Beam 3D Modeling module in RayPhysics/RayPlan Physics. The user must\nbe aware to avoid incorrect dose calculations during treatment planning.",
      "applies": "this device",
      "via": "K180379"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K180379",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "MUJ",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 892.5050",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Radiology",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}