# Raystation

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/raystation-muj>
- Manufacturer: [Raysearch Laboratories AB](https://devicerts.com/manufacturers/raysearch-laboratories-ab) (Sweden, FDA 510(k) applicant)
- Name in the registry: RAYSTATION
- Device class: Class II (US FDA)
- FDA product code: MUJ

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K141860 · decision 2014-10-23 · decision SESE · status: Active

## Recalls

- Z-1195-2025 · Class II · ongoing · initiated 2025-01-16: Inconsistency in the use of Density uncertainty in the RayStation functions Robust optimization, Robust evaluation and Compute perturbed dose for proton and light ion treatment plans when a HU-to-mass density CT calibration curve used in RayStation version 4.5 to 2024B including some service packs.
- Z-1712-2016 · Class II · terminated · initiated 2016-02-10: A software issue with editing tools that use the left mouse button held down, for drawing in and interacting with the patient views and beams eye views. The views can become unsynchronized with the stored data if simultaneously right clicking, pressing Ctrl-S or Ctrl-Z while the left mouse button is held down. This bug does not affect dose computations, which are based on the stored system state.
- Z-2785-2015 · Class II · terminated · initiated 2015-05-29: An error caused by a floating point precision problem tends to cause a problem with the display of isodose lines for dose and dose difference and color tables for dose, dose difference and PET images
- Z-2457-2015 · Class II · terminated · initiated 2015-05-27: Issue with the algorithm for ROI contraction in RayStation 4.7 when non-uniform contraction distances are used. The ROI contraction tool uses six distances as input: right/left, inferior/superior and posterior/anterior. These contraction distances are, in error, pairwise reversed, i.e. right is interpreted as left, inferior is interpreted as superior, and posterior is interpreted as anterior.
- Z-1310-2015 · Class II · terminated · initiated 2015-03-04: An issue with photon dose calculation for DMLC (Dynamic MLC) plans for machines where the MLC is positioned above the jaws, e.g. some Elekta linacs. The magnitude of the error depends on the beam model output factor corrections and on the individual DMLC plan characteristics.

## Identifiers

- 510(k) Number: K141860 [FDA 510(k)]
- FDA product code: MUJ [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 892.5050 [FDA 510(k)]
- Advisory committee: Radiology [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: Raystation, Raysearch Laboratories AB. deviCERTS medical device catalogue. https://devicerts.com/devices/raystation-muj (data as of 2026-11-20).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-20
- JSON: <https://devicerts.com/devices/raystation-muj.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/raystation-muj
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.