  "format": "devicerts.device.v1",
  "cite_as": "Regenerx Ringloc + Modular Acetabular Shells, Biomet, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/regenerx-ringloc-modular-acetabular-shells-lph (data as of 2026-11-21).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/regenerx-ringloc-modular-acetabular-shells-lph",
    "json_url": "https://devicerts.com/devices/regenerx-ringloc-modular-acetabular-shells-lph.json",
    "markdown_url": "https://devicerts.com/devices/regenerx-ringloc-modular-acetabular-shells-lph.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-21",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/regenerx-ringloc-modular-acetabular-shells-lph",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Regenerx Ringloc + Modular Acetabular Shells",
    "registry_name": "REGENERX RINGLOC + MODULAR ACETABULAR SHELLS",
    "url": "https://devicerts.com/devices/regenerx-ringloc-modular-acetabular-shells-lph",
    "manufacturer": {
      "name": "Biomet, Inc.",
      "url": "https://devicerts.com/manufacturers/biomet-inc",
      "country": "United States",
      "country_basis": "FDA 510(k) applicant",
      "fda_registration_country": null
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K070369",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "LPH"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K070369",
      "number_label": null,
      "device_udi_di": null,
      "date": "2007-03-09",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-2463-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-14",
      "reason": "The firm has initiated this action following an investigation which found that the locking ring may  be assembled incorrectly in the acetabular shell.  If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.",
      "applies": "this device",
      "via": "K070369"
    },
    {
      "number": "Z-2452-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-14",
      "reason": "The firm has initiated this action following an investigation which found that the locking ring may  be assembled incorrectly in the acetabular shell.  If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.",
      "applies": "this device",
      "via": "K070369"
    },
    {
      "number": "Z-2449-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-14",
      "reason": "The firm has initiated this action following an investigation which found that the locking ring may  be assembled incorrectly in the acetabular shell.  If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.",
      "applies": "this device",
      "via": "K070369"
    },
    {
      "number": "Z-2465-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-14",
      "reason": "The firm has initiated this action following an investigation which found that the locking ring may  be assembled incorrectly in the acetabular shell.  If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.",
      "applies": "this device",
      "via": "K070369"
    },
    {
      "number": "Z-2450-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-14",
      "reason": "The firm has initiated this action following an investigation which found that the locking ring may  be assembled incorrectly in the acetabular shell.  If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.",
      "applies": "this device",
      "via": "K070369"
    },
    {
      "number": "Z-2461-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-14",
      "reason": "The firm has initiated this action following an investigation which found that the locking ring may  be assembled incorrectly in the acetabular shell.  If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.",
      "applies": "this device",
      "via": "K070369"
    },
    {
      "number": "Z-2455-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-14",
      "reason": "The firm has initiated this action following an investigation which found that the locking ring may  be assembled incorrectly in the acetabular shell.  If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.",
      "applies": "this device",
      "via": "K070369"
    },
    {
      "number": "Z-2457-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-14",
      "reason": "The firm has initiated this action following an investigation which found that the locking ring may  be assembled incorrectly in the acetabular shell.  If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.",
      "applies": "this device",
      "via": "K070369"
    },
    {
      "number": "Z-2454-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-14",
      "reason": "The firm has initiated this action following an investigation which found that the locking ring may  be assembled incorrectly in the acetabular shell.  If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.",
      "applies": "this device",
      "via": "K070369"
    },
    {
      "number": "Z-2451-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-14",
      "reason": "The firm has initiated this action following an investigation which found that the locking ring may  be assembled incorrectly in the acetabular shell.  If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.",
      "applies": "this device",
      "via": "K070369"
    },
    {
      "number": "Z-2462-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-14",
      "reason": "The firm has initiated this action following an investigation which found that the locking ring may  be assembled incorrectly in the acetabular shell.  If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.",
      "applies": "this device",
      "via": "K070369"
    },
    {
      "number": "Z-2458-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-14",
      "reason": "The firm has initiated this action following an investigation which found that the locking ring may  be assembled incorrectly in the acetabular shell.  If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.",
      "applies": "this device",
      "via": "K070369"
    },
    {
      "number": "Z-2466-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-14",
      "reason": "The firm has initiated this action following an investigation which found that the locking ring may  be assembled incorrectly in the acetabular shell.  If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.",
      "applies": "this device",
      "via": "K070369"
    },
    {
      "number": "Z-2459-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-14",
      "reason": "The firm has initiated this action following an investigation which found that the locking ring may  be assembled incorrectly in the acetabular shell.  If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.",
      "applies": "this device",
      "via": "K070369"
    },
    {
      "number": "Z-2448-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-14",
      "reason": "The firm has initiated this action following an investigation which found that the locking ring may  be assembled incorrectly in the acetabular shell.  If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.",
      "applies": "this device",
      "via": "K070369"
    },
    {
      "number": "Z-2460-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-14",
      "reason": "The firm has initiated this action following an investigation which found that the locking ring may  be assembled incorrectly in the acetabular shell.  If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.",
      "applies": "this device",
      "via": "K070369"
    },
    {
      "number": "Z-2464-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-14",
      "reason": "The firm has initiated this action following an investigation which found that the locking ring may  be assembled incorrectly in the acetabular shell.  If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.",
      "applies": "this device",
      "via": "K070369"
    },
    {
      "number": "Z-2456-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-14",
      "reason": "The firm has initiated this action following an investigation which found that the locking ring may  be assembled incorrectly in the acetabular shell.  If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.",
      "applies": "this device",
      "via": "K070369"
    },
    {
      "number": "Z-2453-2011",
      "classification": "",
      "status": "terminated",
      "initiated": "2011-04-14",
      "reason": "The firm has initiated this action following an investigation which found that the locking ring may  be assembled incorrectly in the acetabular shell.  If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.",
      "applies": "this device",
      "via": "K070369"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K070369",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "LPH",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 888.3358",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Orthopedic",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}