Release 2 of Philips Integris Series (V5000, BV5000, H5000, BH5000, HM2000) Systems

by Philips Medical Systems North America, Inc.

510(k) K984545

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

17 closed

Documents

Trust

Unverified

Registry data, not yet checked by deviCERTS. Source FDA 510(k), synced 22 Oct 2026. What this means

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK984545 Date25 Feb 1999ClearedStatusActive

Recalls

17 recalls · FDA

  • 2 recalls, same reason

    • Terminated Z-1140-2019 FDA Class II initiated 19 Feb 2019
    • Terminated Z-1141-2019 FDA Class II initiated 19 Feb 2019

    Additional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or continuous inability of making live fluoroscopy images or exposures.

    • Terminated Z-1280-2016 FDA Class II initiated 24 Sep 2014

    Loss of key image functionality due to a bent pedal of the Footswitch.

    • Terminated Z-1655-2011 FDA Unclassified (pre-2012) initiated 7 Oct 2010

    A wire in the connector of the Generator Grid Switch Supervisor box might short-circuit between 12V power and ground, and it will cause a System breakdown.

    • Terminated Z-2448-2010 FDA Unclassified (pre-2012) initiated 4 Feb 2010

    A lock-up might occur inside the Velara Generator due to intermittent shorts inside the system. The problem could only resolved by a warm or cold system restart. In exceptional cases the power converters of the KVmA unit in the generator are damaged and the system can not be restarted.

    • Terminated Z-2449-2010 FDA Unclassified (pre-2012) initiated 4 Feb 2010

    A lock-up might occur inside the Velara Generator due to intermittent shorts inside the system. The problem can only resolved by a warm or cold system restart. In exceptional cases the power converters of the KVmA unit in the generator are damaged and the system can not be restarted.

  • 4 recalls, same reason

    • Terminated Z-0348-2009 FDA Unclassified (pre-2012) initiated 10 Dec 2008
    • Terminated Z-0345-2009 FDA Unclassified (pre-2012) initiated 10 Dec 2008
    • Terminated Z-0346-2009 FDA Unclassified (pre-2012) initiated 10 Dec 2008
    • Terminated Z-0344-2009 FDA Unclassified (pre-2012) initiated 10 Dec 2008

    Tube failure occurs involving fluoroscopy of large patients and a partial failure of the grid switch.

  • 4 recalls, same reason

    • Terminated Z-2472-2010 FDA Unclassified (pre-2012) initiated 23 Oct 2008
    • Terminated Z-2473-2010 FDA Unclassified (pre-2012) initiated 23 Oct 2008
    • Terminated Z-2471-2010 FDA Unclassified (pre-2012) initiated 23 Oct 2008
    • Terminated Z-2475-2010 FDA Unclassified (pre-2012) initiated 23 Oct 2008

    A high voltage reliability issue resulting in the loss of the x-ray generation and this can cause an intermittent stop of fluoro acquisition and exposure during an examination without warning and hence no image on the display.

    • Terminated Z-0367-2007 FDA Unclassified (pre-2012) initiated 14 Dec 2006

    Movement of the AD5 table or Lateral C-arm (LARC). The table and/or LARC will not move upon request, will move slowly when movement is requested or will not stop moving when requested.

  • 2 recalls, same reason

    • Terminated Z-0207-06 FDA Unclassified (pre-2012) initiated 12 Oct 2005
    • Terminated Z-0206-06 FDA Unclassified (pre-2012) initiated 12 Oct 2005

    Potential for high voltage exposure involving the injector interface connector of patient support tables.

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA Product Code
IZI
Device Class
Class II
Regulation
21 CFR Part 892.1600
Advisory Committee
Radiology
FDA Decision
SESE Substantially Equivalent