Release 2 of Philips Integris Series (V5000, BV5000, H5000, BH5000, HM2000) Systems
by Philips Medical Systems North America, Inc.
510(k) K984545
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK984545 | Date25 Feb 1999 | StatusActive |
Recalls
-
- Terminated Z-1140-2019
- Terminated Z-1141-2019
Additional units have been identified for previous recall Z-1280-2016 (RES 69339): If a Footswitch is frequently used on a non-flat surface or in the pedestal, the footswitch pedals may get bent. This might cause an intermittent or continuous inability of making live fluoroscopy images or exposures.
-
- Terminated Z-1280-2016
Loss of key image functionality due to a bent pedal of the Footswitch.
-
- Terminated Z-1655-2011
A wire in the connector of the Generator Grid Switch Supervisor box might short-circuit between 12V power and ground, and it will cause a System breakdown.
-
- Terminated Z-2448-2010
A lock-up might occur inside the Velara Generator due to intermittent shorts inside the system. The problem could only resolved by a warm or cold system restart. In exceptional cases the power converters of the KVmA unit in the generator are damaged and the system can not be restarted.
-
- Terminated Z-2449-2010
A lock-up might occur inside the Velara Generator due to intermittent shorts inside the system. The problem can only resolved by a warm or cold system restart. In exceptional cases the power converters of the KVmA unit in the generator are damaged and the system can not be restarted.
-
- Terminated Z-0348-2009
- Terminated Z-0345-2009
- Terminated Z-0346-2009
- Terminated Z-0344-2009
Tube failure occurs involving fluoroscopy of large patients and a partial failure of the grid switch.
-
- Terminated Z-2472-2010
- Terminated Z-2473-2010
- Terminated Z-2471-2010
- Terminated Z-2475-2010
A high voltage reliability issue resulting in the loss of the x-ray generation and this can cause an intermittent stop of fluoro acquisition and exposure during an examination without warning and hence no image on the display.
-
- Terminated Z-0367-2007
Movement of the AD5 table or Lateral C-arm (LARC). The table and/or LARC will not move upon request, will move slowly when movement is requested or will not stop moving when requested.
-
- Terminated Z-0207-06
- Terminated Z-0206-06
Potential for high voltage exposure involving the injector interface connector of patient support tables.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- IZI
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1600
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent