Skip to content

revogene

by Genepoc, Inc.

510(k) K170558

revogene is a medical device (FDA Class II; FDA product code OOI, Real Time Nucleic Acid Amplification System) made by Genepoc, Inc., cleared under FDA 510(k) K170558 on 25 May 2017, active as of 22 Oct 2026. 2 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

2 ongoing

This device is named in the recall

Documents

Trust

Unverified

Source FDA 510(k), synced 22 Oct 2026.

Regulatory status

as of registry sync 22 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK170558 Date25 May 2017ClearedStatusActive

Official records

From the FDA record

Received by the FDA
24 Feb 2017
FDA decision
25 May 2017 (Substantially Equivalent)
FDA review time
90 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K170558 (openFDA), as read on 22 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

2 recalls · 2 ongoing · FDA

    • Ongoing Z-1479-2022 FDA Class II initiated 22 Jun 2022

    When a run is aborted, the cooling period protection does not occur and the user is able to open the lid prior to the completion of the cooling period, exposing the user to heated instrument components. Incidental contact with heated instrument components may result in an injury such as minor burns.

    • Ongoing Z-0836-2022 FDA Class II initiated 15 Feb 2022

    The photomultiplier tube, may have compromised functionality in some instruments. As a result, the user is likely to observe an increase in false-positive or indeterminate results.

Timeline

3 dated events

  1. Cleared FDA 510(k) K170558
  2. Recall initiated FDA recalls Z-0836-2022
  3. Recall initiated FDA recalls Z-1479-2022

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
OOI
Device class
Class II
Regulation
21 CFR Part 862.2570
Advisory committee
Clinical Chemistry
FDA decision
SESE Substantially Equivalent

Devices cleared under K170558

2 devices · FDA GUDID

Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.

Devices cleared under K170558
UDI-DISize and differences
Revogene 610210
Meridian Bioscience Inc.
610210 / 61021000840733102318 —On the market
revogene™ 130840
Genepoc, Inc.
130840 / 13084000628055344001 —On the market

Similar devices

Other devices filed under FDA product code OOI (Real Time Nucleic Acid Amplification System).

Other devices from Genepoc, Inc.

  • revogene™ 130840 Open recall

    Class II
  • GenePOC CDiff Open recall

    Class II
  • GenePOC Carba

    Class II
  • GenePOC GBS LB

    Class I

See all devices from Genepoc, Inc. →