# revogene

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/revogene-ooi>
- Manufacturer: [Genepoc, Inc.](https://devicerts.com/manufacturers/genepoc-inc) (Canada, FDA 510(k) applicant)
- Device class: Class II (US FDA)
- FDA product code: OOI

## Registrations and authorizations

- **510(k)** · FDA 510(k) · United States · K170558 · decision 2017-05-25 · decision SESE · status: Active

## Recalls

- Z-1479-2022 · Class II · ongoing · initiated 2022-06-22: When a run is aborted, the cooling period protection does not occur and the user is able to open the lid prior to the completion of the cooling period, exposing the user to heated instrument components. Incidental contact with heated instrument components may result in an injury such as minor burns.
- Z-0836-2022 · Class II · ongoing · initiated 2022-02-15: The photomultiplier tube, may have compromised functionality in some instruments. As a result, the user is likely to observe an increase in false-positive or indeterminate results.

## Identifiers

- 510(k) Number: K170558 [FDA 510(k)]
- FDA product code: OOI [FDA 510(k)]

## Classification

- Device class: Class II [FDA 510(k)]
- Regulation: 21 CFR Part 862.2570 [FDA 510(k)]
- Advisory committee: Clinical Chemistry [FDA 510(k)]
- FDA decision: SESE [FDA 510(k)]

## Source and citation

- Cite as: revogene, Genepoc, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/revogene-ooi (data as of 2026-11-21).
- Data sources: [FDA 510(k)](https://open.fda.gov/apis/device/510k/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-22
- Statuses computed as of: 2026-11-21
- JSON: <https://devicerts.com/devices/revogene-ooi.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/revogene-ooi
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.