> Active: European Union and United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/revolution-ascend-revolution-ascend>
- Manufacturer: [GE Healthcare Japan Corporation](https://devicerts.com/manufacturers/fda-fei-3002808529) (Japan)
- Device class: Class IIb (EU MDR); Class II (US FDA)
- FDA product code: JAK
- EMDN: Z11030606 COMPUTED TOMOGRAPHS - 128 LAYERS OR MORE BUT LESS THAN 256 LAYERS
- GMDN: Full-body CT system (37618)

## Registrations and authorizations

- **EUDAMED** · CE/MDR · European Union · Basic UDI-DI 8406821BUG00117H2 · UDI-DI 00840682146173 · registered 2022-05-17 · status: Active
- **GUDID** · FDA 510(k) · United States · K213938 · issued 2022-02-04 · status: Active · from GUDID
- **GUDID** · GUDID listing · United States · UDI-DI 00840682146173 · listed 2020-11-26 · status: Listed (GUDID: in commercial distribution)
- **GUDID** · FDA 510(k) · United States · K203169 · issued 2020-11-20 · status: Active · from GUDID

## Recalls

- Z-2149-2026 · Class II · ongoing · initiated 2026-03-26: GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
- Z-0605-2025 · Class II · ongoing · initiated 2024-11-05: GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of \~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
- Z-1511-2022 · Class II · ongoing · initiated 2022-06-02: As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place resulting in injury to service personnel.
- Z-2023-2021 · Class II · ongoing · initiated 2021-06-30: When adding one or more intended scan groups names must contain only letters, numbers (0-9), or a space or the user interface will not update to confirm the intended group(s) has been added to the acquisition.

## Identity

- Trade name: Revolution Ascend [EUDAMED]; Revolution [FDA GUDID (AccessGUDID)]
- Generic name: Full-body CT system [FDA GUDID (AccessGUDID)]
- Model / Ref / Catalog: Revolution Ascend [EUDAMED]; Revolution Ascend Elite [EUDAMED]; Revolution Ascend Plus [EUDAMED]; Revolution Ascend Select [EUDAMED]; Ascend [FDA GUDID (AccessGUDID)]; RevolutionAscend [FDA GUDID (AccessGUDID)]

## Identifiers

- Primary DI / UDI-DI: 00840682146173 [EUDAMED, FDA GUDID (AccessGUDID)]
- Basic UDI-DI: 8406821BUG00117H2 [EUDAMED]
- 510(k) Number: K203169 [FDA GUDID (AccessGUDID)]; K213938 [FDA GUDID (AccessGUDID)]
- FDA product code: JAK [FDA GUDID (AccessGUDID)]
- EMDN code: Z11030606 [EUDAMED]
- GMDN term: Full-body CT system (37618) [FDA GUDID (AccessGUDID)]

## Classification

- Device class: Class IIb [EUDAMED]; Class II [FDA GUDID (AccessGUDID)]
- Regulation number: 892.1750 [FDA GUDID (AccessGUDID)]
- EU registration type: Standard device [EUDAMED]
- EU market of registration: France [EUDAMED]
- GUDID distribution status: In Commercial Distribution [FDA GUDID (AccessGUDID)]

## Properties

- Measuring function: Yes [EUDAMED]
- Prescription use: Yes [FDA GUDID (AccessGUDID)]
- Serial number: Yes [FDA GUDID (AccessGUDID)]
- Supplied sterile: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Sterilise before use: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Single use: No [FDA GUDID (AccessGUDID)]
- Implantable: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Reusable: No [EUDAMED]
- Contains latex: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Reprocessed single-use device: No [EUDAMED]
- Contains CMR substances (carcinogenic, mutagenic or toxic to reproduction): No [EUDAMED]
- Incorporates a substance derived from human blood or plasma: No [EUDAMED]
- Contains human tissues or cells: No [EUDAMED]
- Contains animal tissues or cells: No [EUDAMED]
- Contains endocrine-disrupting substances: No [EUDAMED]
- Administering medicine: No [EUDAMED]
- Incorporates a medicinal substance: No [EUDAMED]
- Over-the-counter: No [FDA GUDID (AccessGUDID)]
- Convenience kit: No [FDA GUDID (AccessGUDID)]
- Combination product: No [FDA GUDID (AccessGUDID)]
- HCT/P: No [FDA GUDID (AccessGUDID)]
- Premarket exempt: No [FDA GUDID (AccessGUDID)]
- Lot / batch number: No [FDA GUDID (AccessGUDID)]
- Expiration date: No [FDA GUDID (AccessGUDID)]
- Manufacturing date: No [FDA GUDID (AccessGUDID)]

## Source and citation

- Cite as: Revolution Ascend, GE Healthcare Japan Corporation. deviCERTS medical device catalogue. https://devicerts.com/devices/revolution-ascend-revolution-ascend (data as of 2026-11-26).
- Data sources: [EUDAMED](https://ec.europa.eu/tools/eudamed), [FDA GUDID (AccessGUDID)](https://accessgudid.nlm.nih.gov/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-24
- Statuses computed as of: 2026-11-26
- JSON: <https://devicerts.com/devices/revolution-ascend-revolution-ascend.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/revolution-ascend-revolution-ascend
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.