Revolution CT with Apex edition
by GE Medical Systems, LLC EUDAMEDGUDID
Active
Ref Revolution CT with Apex editionModel Apex Edition
Regulatory status
Recall ongoing. 12 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | Number8406821BUG00154H8 | Date16 Oct 2024 | StatusActive |
| United States | GUDID listing GUDID | Number00840682146319 | Date8 Aug 2019 | StatusIn distribution |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
- European Union
- United States
Recalls
-
- Ongoing Z-2154-2026
- Ongoing Z-2152-2026
- Ongoing Z-2153-2026
- Ongoing Z-2150-2026
GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with certain Revolution Apex, Revolution Ascend, and Revolution CT systems.
-
- Ongoing Z-1297-2025
- Ongoing Z-1300-2025
- Ongoing Z-1303-2025
GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.
-
- Ongoing Z-0242-2022
The accumulated dose is incorrectly displayed in certain situations. Thisoccurs during an interventional exam using the SmartStep option if the table height is adjusted after the exam starts. In this situation, the SmartStep display will show a value that is higher (up to 100 times) than the actual dose delivered.
-
- Ongoing Z-0815-2022
- Ongoing Z-0816-2022
The accumulated dose is incorrectly displayed in certain situations. There are two situations that will result in the SmartStep displayed dose being inaccurately high.
-
- Ongoing Z-2021-2021
- Ongoing Z-2022-2021
When adding one or more intended scan groups names must contain only letters, numbers (0-9), or a space or the user interface will not update to confirm the intended group(s) has been added to the acquisition.
-
- Completed Z-2004-2021
GE Healthcare has become aware of a potential issue on the Revolution CT system where a collimator blade does not find its appropriate home position. This issue could result in the delivered dose of x-ray radiation being either larger or smaller than its prescribed value.
-
- Terminated Z-1129-2021
- Terminated Z-1130-2021
GE Healthcare has become aware of a potential issue on Revolution Apex, Revolution CT with Apex Edition, Revolution CT, and Revolution CT ES systems where the table pinch protector could be damaged or missing leading to exposed table screws which could cause a laceration due to a sharp edge.
-
- Terminated Z-0694-2021
- Terminated Z-0695-2021
A potential for protocol(s) with unintended patient orientation or scan setting when using Intelligent Protocoling
-
- Terminated Z-0306-2021
There is a potential for a smudge artifact that could be suspect for pathology in some images due to incorrect settings.
Specification
- Trade Name
- Revolution CT with Apex edition EUDAMEDRevolution CT GUDID
- Generic Name
- Full-body CT system
- Model / Ref / Catalog
- Apex Edition EUDAMEDRevolution CT with Apex edition EUDAMEDwith Apex edition GUDID5826588 GUDID
- Primary DI / UDI-DI
- 00840682146319
- 510(k) Number
- FDA Product Code
- JAK
- GMDN Term
- Full-body CT system
- Regulation Number
- 892.1750
- Market of Registration
- Denmark
- Commercial Distribution
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Model / Reference | Apex Edition Revolution CT with Apex edition | with Apex edition |
| Device Class | Class IIb | Class II |
Manufacturer details
- EUDAMED SRN
- US-MF-000018315
- Authorised representative
- GE Medical Systems SCS FR-AR-000000344
Other devices from GE Medical Systems, LLC
Properties & raw data
Show properties, registry sources and raw records
Other properties (21)
- Sterile: No
- Single Use: No
- Implantable: No
- Reusable: No
- Contains Latex: No
- Reprocessed: No
- CMR Substance: No
- Human-derived: No
- Human Tissues: No
- Animal Tissues: No
- Endocrine Disruptor: No
- Administering Medicine: No
- Incorporates Medicinal Product: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No
- Manufacturing Date: No
Sources (2)
- EUDAMED 4bf91ebd-bf95-41be-887a-6877098014ce
- FDA UDI a534801f-4fb1-4500-93b5-689d6f13c91a