Revolution External Plating System
by New Standard Device Dba Metalogix 2 variants
510(k) K181630
Active
510(k)
Valid until
No expiry
FDA recalls
3 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK181630 | Date29 May 2019 | StatusActive |
Recalls
-
- Ongoing Z-1649-2022
Complaints were received for Self-drill Half-Pin tip breakages occurring during surgery.
-
- Ongoing Z-1280-2022
Due to manufacturing processing error, wrench causing stripping when torqueing nuts.
-
- Ongoing Z-1824-2023
The threads on posts stripping out when tensioned during application process.
-
- Completed Z-1236-2022
1.8mm truss wires are being removed to facilitate the introduction of 2.0mm truss wires. The increased wire diameter and new manufacturing process are meant to provide a stronger wire and reduced breakage.
-
- Terminated Z-0318-2020
Non-highly polished implants may adhere to the LDPE bag (previous packaging configuration).
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- KTT
- Device Class
- Class II
- Regulation
- 21 CFR Part 888.3030
- Advisory Committee
- Orthopedic
- FDA Decision
- SESE Substantially Equivalent