# Ringloc + Hip System 106026

> Active: United States. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/ringloc-hip-system-106026>
- Manufacturer: [Biomet Orthopedics, Inc.](https://devicerts.com/manufacturers/biomet-orthopedics-inc) (United States)
- Name in the registry: RINGLOC + HIP SYSTEM 106026
- Device class: Class II (US FDA)
- FDA product code: MBL, MAY, JDG, KWZ, MEH, LWJ, JDI, LZO, LPH
- GMDN: Acetabular shell (43167)

## Registrations and authorizations

- **GUDID** · GUDID listing · United States · UDI-DI 00880304455030 · listed 2015-10-24 · status: Listed (GUDID: in commercial distribution)
- **GUDID** · FDA 510(k) · United States · K070369 · decision 2007-03-09 · decision SESE · status: Active · from GUDID

## Recalls

- Z-2463-2011 ·  · terminated · initiated 2011-04-14: The firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.
- Z-2452-2011 ·  · terminated · initiated 2011-04-14: The firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.
- Z-2449-2011 ·  · terminated · initiated 2011-04-14: The firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.
- Z-2465-2011 ·  · terminated · initiated 2011-04-14: The firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.
- Z-2450-2011 ·  · terminated · initiated 2011-04-14: The firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.
- Z-2461-2011 ·  · terminated · initiated 2011-04-14: The firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.
- Z-2455-2011 ·  · terminated · initiated 2011-04-14: The firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.
- Z-2457-2011 ·  · terminated · initiated 2011-04-14: The firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.
- Z-2454-2011 ·  · terminated · initiated 2011-04-14: The firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.
- Z-2451-2011 ·  · terminated · initiated 2011-04-14: The firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.
- Z-2462-2011 ·  · terminated · initiated 2011-04-14: The firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.
- Z-2458-2011 ·  · terminated · initiated 2011-04-14: The firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.
- Z-2466-2011 ·  · terminated · initiated 2011-04-14: The firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.
- Z-2459-2011 ·  · terminated · initiated 2011-04-14: The firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.
- Z-2448-2011 ·  · terminated · initiated 2011-04-14: The firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.
- Z-2460-2011 ·  · terminated · initiated 2011-04-14: The firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.
- Z-2464-2011 ·  · terminated · initiated 2011-04-14: The firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.
- Z-2456-2011 ·  · terminated · initiated 2011-04-14: The firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.
- Z-2453-2011 ·  · terminated · initiated 2011-04-14: The firm has initiated this action following an investigation which found that the locking ring may be assembled incorrectly in the acetabular shell. If the locking ring is incorrectly assembled in the acetabular shell, the liner cannot be completely seated in the shell.

## Identity

- Trade name: RINGLOC + HIP SYSTEM [FDA GUDID (AccessGUDID)]
- Generic name: Acetabular shell [FDA GUDID (AccessGUDID)]
- Model / Ref / Catalog: 106026 [FDA GUDID (AccessGUDID)]

## Identifiers

- Primary DI / UDI-DI: 00880304455030 [FDA GUDID (AccessGUDID)]
- 510(k) Number: K070369 [FDA GUDID (AccessGUDID)]
- FDA product code: MBL [FDA GUDID (AccessGUDID)]; MAY [FDA GUDID (AccessGUDID)]; JDG [FDA GUDID (AccessGUDID)]; KWZ [FDA GUDID (AccessGUDID)]; MEH [FDA GUDID (AccessGUDID)]; LWJ [FDA GUDID (AccessGUDID)]; JDI [FDA GUDID (AccessGUDID)]; LZO [FDA GUDID (AccessGUDID)]; LPH [FDA GUDID (AccessGUDID)]
- GMDN term: Acetabular shell (43167) [FDA GUDID (AccessGUDID)]

## Classification

- Device class: Class II [FDA GUDID (AccessGUDID)]
- Regulation number: 888.3358 [FDA GUDID (AccessGUDID)]; 888.3353 [FDA GUDID (AccessGUDID)]; 888.3360 [FDA GUDID (AccessGUDID)]; 888.3310 [FDA GUDID (AccessGUDID)]; 888.3350 [FDA GUDID (AccessGUDID)]
- GUDID distribution status: In Commercial Distribution [FDA GUDID (AccessGUDID)]

## Properties

- Supplied sterile: Yes [FDA GUDID (AccessGUDID)]
- Single use: Yes [FDA GUDID (AccessGUDID)]
- Implantable: Yes [FDA GUDID (AccessGUDID)]
- Prescription use: Yes [FDA GUDID (AccessGUDID)]
- Lot / batch number: Yes [FDA GUDID (AccessGUDID)]
- Expiration date: Yes [FDA GUDID (AccessGUDID)]
- Sterilise before use: No [FDA GUDID (AccessGUDID)]
- Contains latex: No [FDA GUDID (AccessGUDID)]
- Over-the-counter: No [FDA GUDID (AccessGUDID)]
- Convenience kit: No [FDA GUDID (AccessGUDID)]
- Combination product: No [FDA GUDID (AccessGUDID)]
- HCT/P: No [FDA GUDID (AccessGUDID)]
- Premarket exempt: No [FDA GUDID (AccessGUDID)]
- Serial number: No [FDA GUDID (AccessGUDID)]
- Manufacturing date: No [FDA GUDID (AccessGUDID)]

## Source and citation

- Cite as: Ringloc + Hip System 106026, Biomet Orthopedics, Inc. deviCERTS medical device catalogue. https://devicerts.com/devices/ringloc-hip-system-106026 (data as of 2026-11-21).
- Data sources: [FDA GUDID (AccessGUDID)](https://accessgudid.nlm.nih.gov/), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-10-23
- Statuses computed as of: 2026-11-21
- JSON: <https://devicerts.com/devices/ringloc-hip-system-106026.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/ringloc-hip-system-106026
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.