Skip to content

Roche Cobas Ibtegra Reagent Cassettes

by Roche Diagnostic Systems, Inc. 510(k)

Active United States

510(k) K964457

Regulatory status

as of 29 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK964457 Date13 Jan 1997ClearedStatusActive

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Recalls

2 recalls · FDA

    • Terminated Z-1146-2019 FDA Class II initiated 17 Dec 2018 on shared 510(k) K964457

    New endogenous interference claims have been established for the following assays used on the cobas c 311 analyzer and the cobas c 501, 502, 701, 702, and Modular Analytics P/D modules which may have potential medical risk. ***Update: Cobas c111 and COBAS INTEGRA 400 plus analyzers are also impacted for BILT3 and LACT2 assays***

    • Terminated Z-1926-2015 FDA Class II initiated 7 May 2015 on shared 510(k) K964457

    Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Specification

510(k)

510(k) Number
FDA Product Code
JQB
Device Class
Class I
Regulation
21 CFR Part 862.1360
Advisory Committee
Clinical Chemistry
FDA Decision
SESE Substantially Equivalent

Other devices from Roche Diagnostic Systems, Inc.