Roche Cobas Ibtegra Reagent Cassettes
by Roche Diagnostic Systems, Inc. 510(k)
Active
510(k) K964457
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK964457 | Date13 Jan 1997 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-1146-2019
New endogenous interference claims have been established for the following assays used on the cobas c 311 analyzer and the cobas c 501, 502, 701, 702, and Modular Analytics P/D modules which may have potential medical risk. ***Update: Cobas c111 and COBAS INTEGRA 400 plus analyzers are also impacted for BILT3 and LACT2 assays***
-
- Terminated Z-1926-2015
Possible drug interference when using the Trinder reaction method on multiple Cobas Analyzers. Results of Trinder tests may be falsely decreased when the tests are run with urine samples (Creatinine plus and Uric Acid applications only) or blood samples collected before sufficient time has lapsed after the administration of the drugs Acetaminophen and N-acetylcysteine.
Specification
- 510(k) Number
- FDA Product Code
- JQB
- Device Class
- Class I
- Regulation
- 21 CFR Part 862.1360
- Advisory Committee
- Clinical Chemistry
- FDA Decision
- SESE Substantially Equivalent