Rosa Surgical Device, Model Rosa 1.1
by Medtech S.A
510(k) K092239
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK092239 | Date17 Nov 2009 | StatusActive |
Recalls
-
- Terminated Z-3117-2017
Complaints were received reporting the system would freeze/shut down while in Fulgurate mode.
-
- Terminated Z-0160-2018
Potential failure of a force feedback sensor that may result in continued movement of an operating arm (rather than an immediate stop as intended).
-
- Terminated Z-0184-2018
The software issue described was corrected in the modification to the MXTTOUT controller parameter settings.
-
- Terminated Z-0161-2018
Software corrections reactivating the cooperative endoscopy mode.
-
- Terminated Z-0102-2018
An undetected shift between the information displayed in the navigation software and the actual patient anatomy
-
- Terminated Z-0096-2018
Possible break in the connector of the Force Sensor.
-
- Terminated Z-0178-2018
Issue with optional neuro-endoscopy module detected under specific conditions during internal testing.
-
- Terminated Z-0099-2018
Observed instability of the device.
-
- Terminated Z-0100-2018
Potential failure of the optical distance sensor due to cable disconnection.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- HAW
- Device Class
- Class II
- Regulation
- 21 CFR Part 882.4560
- Advisory Committee
- Neurology
- FDA Decision
- SESE Substantially Equivalent