Rotarex Atherectomy System
by Bard Peripheral Vascular, Inc. 510(k)
Active
510(k) K242757
Regulatory status
Recall ongoing. 1 ongoing FDA recall. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK242757 | Date30 Jan 2025 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Ongoing Z-1244-2025
Atherectomy Catheter eIFU updated to clarify and emphasize procedural steps intended to reduce the likelihood of catheter breakage events. Catheter has an outer cylinder connected to a rotating helix, which could fracture and/or break, which would require retrieval, and helix facture/break could cause vessel injury and lead to severe bleeding.
Specification
- 510(k) Number
- FDA Product Code
- MCW
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.4875
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent