  "format": "devicerts.device.v1",
  "cite_as": "Sarn Centrifugal Pump, Terumo Cardiovascular Systems Corporation. deviCERTS medical device catalogue. https://devicerts.com/devices/sarn-centrifugal-pump-kfm (data as of 2026-11-26).",
  "source": {
    "canonical_url": "https://devicerts.com/devices/sarn-centrifugal-pump-kfm",
    "json_url": "https://devicerts.com/devices/sarn-centrifugal-pump-kfm.json",
    "markdown_url": "https://devicerts.com/devices/sarn-centrifugal-pump-kfm.md",
    "publisher": "deviCERTS (devicerts.com)",
    "data_sources": [
      {
        "registry": "FDA 510(k)",
        "url": "https://open.fda.gov/apis/device/510k/"
      },
      {
        "registry": "FDA recalls",
        "url": "https://open.fda.gov/apis/device/recall/"
      }
    ],
    "registry_records_last_synced": "2026-10-22",
    "status_as_of": "2026-11-26",
    "license": "CC BY 4.0",
    "license_url": "https://creativecommons.org/licenses/by/4.0/",
    "license_scope": "The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.",
    "attribution": "DevicErts, https://devicerts.com/devices/sarn-centrifugal-pump-kfm",
    "disclaimer": "Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails."
  },
  "device": {
    "name": "Sarn Centrifugal Pump",
    "registry_name": "SARN CENTRIFUGAL PUMP",
    "url": "https://devicerts.com/devices/sarn-centrifugal-pump-kfm",
    "manufacturer": {
      "name": "Terumo Cardiovascular Systems Corporation",
      "url": "https://devicerts.com/manufacturers/terumo-cardiovascular-systems-corporation",
      "country": null,
      "country_basis": null,
      "fda_registration_country": "United States"
    },
    "status": {
      "state": "active",
      "label": "Active",
      "detail": "United States"
    },
    "classification": [
      {
        "class": "Class II",
        "system": "US FDA"
      }
    ],
    "identifiers": [
      {
        "label": "510(k) Number",
        "values": [
          {
            "value": "K112229",
            "registries": [
              "FDA 510(k)"
            ]
          }
        ]
      }
    ],
    "product_codes": [
      "KFM"
    ],
    "emdn": [],
    "gmdn": [],
    "description": null,
    "noindex": false
  },
  "authorizations": [
    {
      "registry": "510(k)",
      "region": "United States",
      "type": "FDA 510(k)",
      "number": "K112229",
      "number_label": null,
      "device_udi_di": null,
      "date": "2011-12-09",
      "date_kind": "decision",
      "decision_code": "SESE",
      "expires": null,
      "withdrawn": null,
      "status": "Active",
      "status_note": null,
      "derived_from_gudid": false
    }
  ],
  "no_longer_listed_by": [],
  "certificates": [],
  "recalls": [
    {
      "number": "Z-1235-2017",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2015-05-27",
      "reason": "Terumo Cardiovascular Systems Corporation has received complaints indicating that the Sarns Centrifugal Pump (disposable pump heads) have exhibited leaks. The leaks were detected during priming of the bypass circuit and during CPB procedures.\n\nIn each of the reported incidents, there were no reports of patient injury as a result of the centrifugal pump leaks.\n\nTerumo Cardiovascular System's investigation into the incident has identified the cause of the leaks to be cracks in both the Top Housing and Magnet Back Housing of the pump. When the cracks cause the housing to become compromised, it is possible for fluid to leak from the pump head.",
      "applies": "this device",
      "via": "K112229"
    },
    {
      "number": "Z-2028-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2014-04-14",
      "reason": "During set-up and priming of the bypass circuit, leaks were detected on the centrifugal pumphead.",
      "applies": "this device",
      "via": "K112229"
    },
    {
      "number": "Z-2029-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2014-04-14",
      "reason": "During set-up and priming of the bypass circuit, leaks were detected on the centrifugal pumphead.",
      "applies": "this device",
      "via": "K112229"
    },
    {
      "number": "Z-2030-2014",
      "classification": "Class II",
      "status": "terminated",
      "initiated": "2014-04-14",
      "reason": "During set-up and priming of the bypass circuit, leaks were detected on the centrifugal pumphead.",
      "applies": "this device",
      "via": "K112229"
    }
  ],
  "eu_market_availability": null,
  "properties": [
    {
      "group": "Identifiers",
      "fields": [
        {
          "label": "510(k) Number",
          "values": [
            {
              "value": "K112229",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA product code",
          "values": [
            {
              "value": "KFM",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    },
    {
      "group": "Classification",
      "fields": [
        {
          "label": "Device class",
          "values": [
            {
              "value": "Class II",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Regulation",
          "values": [
            {
              "value": "21 CFR Part 870.4360",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "Advisory committee",
          "values": [
            {
              "value": "Cardiovascular",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        },
        {
          "label": "FDA decision",
          "values": [
            {
              "value": "SESE",
              "registries": [
                "FDA 510(k)"
              ]
            }
          ]
        }
      ]
    }
  ]
}