Sarns 8000 Modular Perfusion System
510(k) K915183
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK915183 | Date8 Jul 1992 | StatusActive |
Recalls
-
- Terminated Z-0129-2013
- Terminated Z-0131-2013
- Terminated Z-0128-2013
- Terminated Z-0130-2013
- Terminated Z-0127-2013
- Terminated Z-0126-2013
Terumo Cardiovascular Systems (Terumo CVS) has received multiple reports since 2007 of various malfunctions of the power distribution and battery back-up systems for the Sarns Modular Perfusion System 8000. Very few of the reports occurred during cardiopulmonary bypass; the majority of the malfunctions occurred during set-up or were reported during routine preventative maintenance. The malfunction
-
- Terminated Z-0280-2013
- Terminated Z-0283-2013
- Terminated Z-0285-2013
- Terminated Z-0282-2013
- Terminated Z-0281-2013
- Terminated Z-0284-2013
Terumo Cardiovascular Systems (Terumo CVS) has received multiple reports since 2007 of various malfunctions of the arterial and cardioplegia monitors for the Sarns Modular Perfusion System 8000. The malfunctions can be broadly categorized as follows monitor power malfunctions, inability to measure pressure and inaccurate and/or false pressure alarms. If the malfunction of the arterial or cardio
-
- Terminated Z-2218-2012
During a retrospective review of complaint events, TCVS identified on 06/15/12 that three (3) Sarns" System 8000 bases were manufactured and distributed from 11/13/06-11/29/06 with an incorrect incorrect amperage rating identified on the name plate label. The mislabeling was identified in January 2007 a set up process of the automark label printer. All units were corrected in the field in Marc
-
- Terminated Z-2049-2012
- Terminated Z-2047-2012
- Terminated Z-2045-2012
- Terminated Z-2048-2012
- Terminated Z-2046-2012
- Terminated Z-2044-2012
Terumo Cardiovascular Systems has received several reports of malfunctions of the Air Bubble Detection System used with its heart-lung machines over the last three years. The malfunctions were expressed in various failure modes (over sensitivity, false alarms and were found to have various root causes. The malfunctions occurred in systems that were excluded from prior corrective actions or in syst
-
- Terminated Z-1453-2012
- Terminated Z-1451-2012
- Terminated Z-1452-2012
- Terminated Z-1450-2012
Terumo received multiple reports of malfunctions of the Cardioplegia and arterial monitors for the Sarns Modular Perfusion System 8000. Upon review of complaint investigation, the malfunctions are caused by or could have been caused by soldering anomalies on the circuit board of the monitor. The compromised solder joints could cause intermittent or loss of display, or could cause monitor reset, po
-
- Terminated Z-0303-2012
Terumo Cardiovascular Systems is aware on one instance in which an incorrectly labeled printed circuit board for the Sarns Modular Perfusion System 8000 was shipped to a single customer. The affected board mislabeled two connection between the board and the transformer.
-
- Terminated Z-2613-2011
- Terminated Z-2620-2011
- Terminated Z-2617-2011
- Terminated Z-2614-2011
- Terminated Z-2616-2011
- Terminated Z-2618-2011
- Terminated Z-2621-2011
- Terminated Z-2619-2011
- Terminated Z-2615-2011
The firm has recieved complaints regarding 35 pressure monitoring kits in which the membrane in the kit separated form the pressure dome housing. The firm determined that the defect was caused by a change in the pressure dome's manufacturing process.
-
- Terminated Z-0562-2011
- Terminated Z-0563-2011
The system 8000 roller pumps may fail resulting in a persistent pump stop.
-
- Terminated Z-2348-2012
Terumo Cardiovascular Systems (Terumo CVS) received complaints from our Affiliate, Terumo Dubai that the safety monitors were tripping off. Terumo CVS' investigation determined that 8 monitors were manufactured with an underrated circuit breaker due to an error where an associate pulled an underrated circuit breaker from an incorrect inventory location. Since this time, inventory locations hav
-
- Terminated Z-0241-2007
- Terminated Z-0239-2007
- Terminated Z-0236-2007
- Terminated Z-0237-2007
- Terminated Z-0240-2007
- Terminated Z-0238-2007
The hand crank handle may separate and detach from the unit during use.
-
- Terminated Z-1004-2007
- Terminated Z-1002-2007
- Terminated Z-1003-2007
Roller Pump: Due to a circuit board malfunction the roller pumps may stop during use without displaying an error message, may stop during use with an overspeed error posted on the local display, or may fail to start/restart resulting in the pump(s) being non-functional.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- DTQ
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.4220
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent
Overview
AI-generated summary, 22 Sep 2026
The Sarns 8000 Modular Perfusion System is a heart-lung machine console designed for use in cardiopulmonary bypass procedures. This modular platform features a base that can be configured with four or five pumps, including options for roller or centrifugal pump heads.
The system is equipped with arterial and cardioplegia monitors for tracking temperature, pressure, and time, as well as modules for pulse, communications, and venous occlusion. This device has received FDA 510(k) clearance under K915183.
Other devices from 3M Health Care, Sarns
Questions
AI-generated answers, 22 Sep 2026
What is the primary function of the Sarns 8000 Modular Perfusion System?
This device serves as a heart-lung machine console specifically designed to support cardiopulmonary bypass procedures by managing blood flow and monitoring vital parameters during surgery.
How can the Sarns 8000 system be configured for clinical use?
The system is built on a modular platform that allows for a base configuration of either four or five pumps. Clinicians can select from various pump head types, including roller or centrifugal options, to meet specific procedural requirements.
What monitoring capabilities are integrated into the console?
The system includes dedicated arterial and cardioplegia monitors. These modules are designed to track essential patient data, including temperature, pressure, and time, to assist in the management of the bypass procedure.
What is the regulatory status of this perfusion system?
The Sarns 8000 Modular Perfusion System has received FDA 510(k) clearance under K915183, confirming it has been reviewed and determined to be substantially equivalent to existing devices for its intended use.