Sarns Cardioplegia Cannulae (Var. Types
510(k) K822210
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK822210 | Date24 Aug 1982 | StatusActive |
Recalls
-
- Terminated Z-2555-2011
- Terminated Z-2556-2011
The presence of adhesive and/or skiving of plastic material observed inside the distal tip of the cannula, which could lead to particulate and/or decrease in the internal diameter. The plastic skiving may detach and enter the patient, embolizing the coronary vasculature. The potential risk due to decrease in the internal tip diameter is inadequate delivery of cardioplegia and/or prolonged procedu
-
- Terminated Z-2554-2011
Terumo has discovered several instances of adhesive build-up and or skiving of plastic (plastic flash) inside the distal tip of the Terumo Coronary Ostia Cannula. The defect was discovered during in-house inspection after Terumo CVS implemented a new inspection process for the cannula. There have been no reports of the defect in clinical use.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- DWF
- Device Class
- Class II
- Regulation
- 21 CFR Part 870.4210
- Advisory Committee
- Cardiovascular
- FDA Decision
- SESE Substantially Equivalent