SCENARIA Phase 3 Whole-body X-ray CT System
by Hitachi Medical Systems America, Inc.
510(k) K150595
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK150595 | Date30 Oct 2015 | StatusActive |
Recalls
-
- Terminated Z-0271-2020
There is a possibility that the cables to the controller unit may not be affixed properly, creating the potential for a cable to detach during scanning and eject from the system.
-
- Terminated Z-2839-2018
The firm received a complaint where, during patient positioning, the technologist was adjusting the patient table height and the table failed to stop when the control was released and contacted the CT gantry. The problem could not be duplicated at the site, so to avoid further issues while a root cause investigation was pursued, Hitachi choose to add an optional touch sensor to the gantry to add a redundant level of protection. It is conceivable that a patient s extremity may be caught between the tabletop and the gantry surface during uncontrolled motion or that the table contact of the gantry might cause internal damage to the unit.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- JAK
- Device Class
- Class II
- Regulation
- 21 CFR Part 892.1750
- Advisory Committee
- Radiology
- FDA Decision
- SESE Substantially Equivalent