Scorpio 72-15-0508
Active
510(k)
Valid until
No expiry
FDA recalls
3 ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK261898 | Date6 Aug 2026 | StatusActive |
| United States | GUDID listing | Number07613154020140 | Date24 Sep 2015 | StatusListed |
Recalls
-
- Ongoing Z-3206-2026
- Ongoing Z-3197-2026
Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
-
- Ongoing Z-0883-2024
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- Trade name
- Scorpio
- Generic name
- Tibial insert
- Model / Ref / Catalog
- 72-15-0508
- Primary DI / UDI-DI
- 07613154020140
- FDA product code
- JWH
- GMDN term
- Tibial insert
- Device class
- Class II
- Regulation number
- 888.3560
- GUDID distribution status
- In Commercial Distribution
Other devices from Howmedica Osteonics Corp.
Properties
Other properties (10)
- Sterilise before use: No
- Contains latex: No
- Over-the-counter: No
- Convenience kit: No
- Combination product: No
- HCT/P: No
- Premarket exempt: No
- Serial number: No
- Expiration date: No