Seednet System and Seednetgold System, Models GP6T5, GP5T5, GP5T3, HP5T5 & HP5T3
by Galil Medical Ltd 510(k)
Active
510(k) K021261
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK021261 | Date3 Jul 2002 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
United States — FDA 510(k) Active
Recalls
-
- Terminated Z-1032-2011
This voluntary correction has been initiated due to the discovery of a damaged optocoupler on one of our Presice Systems during periodic preventive maintenance. The optocoupler sends a signal across the locking mechanism in the Presice and Seednet MRI Systems, enabling the software to recognize if the channel on the system manifold is locked or unlocked. If an optocoupler is damaged or malfunction
Specification
- 510(k) Number
- FDA Product Code
- GEH
- FDA applicant
- Galil Medical , Ltd.
- Device Class
- Class II
- Regulation
- 21 CFR Part 878.4350
- Advisory Committee
- General, Plastic Surgery
- FDA Decision
- SESE Substantially Equivalent