by Siemens Medi Cal Solutions, Inc. 510(k)

Active United States Open recall

510(k) K150493

Regulatory status

as of 29 Oct 2026

Recall ongoing. 6 ongoing FDA recalls name this device. See recalls

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK150493 Date30 Jun 2015ClearedStatusActive

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Recalls

11 recalls · 6 ongoing · FDA

    • Ongoing Z-0775-2023 FDA Class II initiated 16 Nov 2022

    The firm will be performing a software update to address a software error which affects the listed products. This correction addresses four potential software issues: 1) "PASSWORD STORE CORRUPTED" error message during system boot; 2) Subsystem crash during examination; 3) Dialog Monitor Computer (DMC) application crash while loading a study; and 4) Software crash due to system internal timeout. Issue 1 may lead to a delay or interruption of procedure. Issues 2, 3, and 4 may result in delay in starting or continuing the examination, and may also prevent the operator from starting or continuing a study;

  • 2 recalls, same reason

    • Ongoing Z-1426-2022 FDA Class II initiated 15 Jun 2022
    • Ongoing Z-1425-2022 FDA Class II initiated 15 Jun 2022

    Sensis/Sensis Vibe SoftwareVD12A, the ComboBox may encounter a disconnect of its communication during the first patient examination of the day or after a longer period of inactivity to the Sensis Vibe system resulting in no vital signs being available.

    • Ongoing Z-1243-2022 FDA Class II initiated 13 Apr 2022

    Siemens has become aware of three potential software issues with AXIOM Sensis or Sensis/ Sensis Lite systems. This may lead to a hazardous situation for patients if treatment cannot be continued on the system and treatment needs to be continued on an alternate system.

    • Ongoing Z-0499-2022 FDA Class II initiated 23 Nov 2021

    Software error which affects Sensis Vibe Hemo, Sensis and Sensis Vibe Combo systems with software version VD12A, which could cause possible hazard to patients, operators, or other persons and equipment. Under certain sporadic circumstances, the CO (Cardiac Output) measurement using the Thermodilution method will temporarily no longer be possible.

    • Ongoing Z-0226-2022 FDA Class II initiated 5 Oct 2021 on shared 510(k) K150493

    System operator manual states that the system should be rebooted once, every 7 days. In some cases, users are not following these instructions and the system is running for more than 7 days. Under rare circumstances, this may result in a partial freeze of the user interface which would no longer update the vital signs

    • Terminated Z-1558-2022 FDA Class III initiated 12 Jul 2022

    The Sensis Signal Input Box provides hooks as a milled item on the bottom of the box support the possibility of mounting the Signal Input Box on an accessory rail (e.g., at the patient table), In the event the mounting position deviates from the default position, the following are possible: Patient touching the Signal Input Box may result in a touch leakage current to the patient; Spillage of liquid entering the Signal Input Box

    • Terminated Z-1434-2021 FDA Class II initiated 4 Mar 2021

    Due to the configuration of certain Windows Service Permissions within the operating systems of the Sensis/ Sensis Vibe computer, there is a risk for exposure of sensitive information, manipulation of data, or Denial of Service attacks and could result in incorrect diagnostic or therapeutic decisions

    • Terminated Z-1173-2021 FDA Class II initiated 27 Jan 2021

    System may sporadically freeze (lock-up) during operation or while being in an idle state, no longer possible to interact with the system

    • Terminated Z-2921-2020 FDA Class II initiated 24 Jul 2020

    The connector linking the EtCO2 module to the HemoBox may break and become damaged, if the connector is re-attached to the HemoBox in this damaged state, under certain incorrect pin alignment conditions, the HemoBox may stop functioning and all vital signs are lost

    • Terminated Z-0936-2019 FDA Class II initiated 29 Nov 2018

    A software error may result in a system crash. The system must be restarted before the clinical procedure can be continued. The ablation treatment must be performed with a different system or without the use of the interface.

Recalls on a shared submission name another device cleared under the same FDA submission as this one.

Specification

510(k)

510(k) Number
FDA Product Code
DQK
Device Class
Class II
Regulation
21 CFR Part 870.1425
Advisory Committee
Cardiovascular
FDA Decision
SESE Substantially Equivalent

Overview

AI-generated summary, 26 Sep 2026

Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data above is the record.

The Sensis is a Class II cardiac pacemaker manufactured by Siemens Medi Cal Solutions, Inc. It is used in cardiovascular applications. The device has been cleared under the FDA 510(k) process. The Sensis is substantially equivalent to other devices in its class.

The Sensis is supplied as a single device and is intended for use in a medical setting. It is a reusable device, but specific storage and handling instructions are not provided. The device has been assigned a product code of DQK and has been cleared under the K-number K150493 in 2015. The FDA has issued a 510(k) clearance for this model.

Other devices from Siemens Medi Cal Solutions, Inc.

See all devices from Siemens Medi Cal Solutions, Inc. →

Questions

AI-generated answers, 26 Sep 2026

Written by a language model from registry data. Not verified by deviCERTS or the manufacturer. The registry data above is the record.

  • What is the Sensis used for?

    The Sensis is used in cardiovascular applications, specifically as a cardiac pacemaker.

  • Is the Sensis a single-use or reusable device?

    The Sensis is a reusable device, but specific storage and handling instructions are not provided.

  • What is the regulatory status of the Sensis?

    The Sensis has been cleared under the FDA 510(k) process, with a K-number of K150493, indicating substantial equivalence to other devices in its class.