# Sensis Vibe Combo 11007642

> Active: European Union. Medical device record on deviCERTS, compiled from public registries.

- Canonical page: <https://devicerts.com/devices/sensis-vibe-combo-11007642>
- Manufacturer: [Siemens Healthcare GmbH](https://devicerts.com/manufacturers/siemens-healthcare-gmbh) (Germany)
- Device class: Class IIb (EU MDR); Class II (US FDA)
- FDA product code: DQK
- EMDN: Z12050701 COMPLETE CARDIOGRAPHY SYSTEMS
- GMDN: Cardiac catheterization laboratory computer (10980)

## Registrations and authorizations

- **EUDAMED** · CE/MDR · European Union · Basic UDI-DI 0405686900144UU · UDI-DI 04056869010205 · registered 2026-06-22 · status: Active
- **GUDID** · GUDID listing · United States · UDI-DI 04056869010205 · listed 2017-02-13 · status: Listed (GUDID: in commercial distribution)

## Recalls

- Z-0775-2023 · Class II · ongoing · initiated 2022-11-16: The firm will be performing a software update to address a software error which affects the listed products. This correction addresses four potential software issues: 1) "PASSWORD STORE CORRUPTED" error message during system boot; 2) Subsystem crash during examination; 3) Dialog Monitor Computer (DMC) application crash while loading a study; and 4) Software crash due to system internal timeout. Issue 1 may lead to a delay or interruption of procedure. Issues 2, 3, and 4 may result in delay in starting or continuing the examination, and may also prevent the operator from starting or continuing a study;
- Z-1558-2022 · Class III · terminated · initiated 2022-07-12: The Sensis Signal Input Box provides hooks as a milled item on the bottom of the box support the possibility of mounting the Signal Input Box on an accessory rail (e.g., at the patient table), In the event the mounting position deviates from the default position, the following are possible: Patient touching the Signal Input Box may result in a touch leakage current to the patient; Spillage of liquid entering the Signal Input Box
- Z-1426-2022 · Class II · ongoing · initiated 2022-06-15: Sensis/Sensis Vibe SoftwareVD12A, the ComboBox may encounter a disconnect of its communication during the first patient examination of the day or after a longer period of inactivity to the Sensis Vibe system resulting in no vital signs being available.
- Z-0499-2022 · Class II · ongoing · initiated 2021-11-23: Software error which affects Sensis Vibe Hemo, Sensis and Sensis Vibe Combo systems with software version VD12A, which could cause possible hazard to patients, operators, or other persons and equipment. Under certain sporadic circumstances, the CO (Cardiac Output) measurement using the Thermodilution method will temporarily no longer be possible.

## Identity

- Trade name: Sensis Vibe Combo [EUDAMED, FDA GUDID (AccessGUDID)]
- Generic name: Cardiac catheterization laboratory computer [FDA GUDID (AccessGUDID)]
- Device name: MDR\_01\_IIb\_FOR\_ACDV\_Sensis Vibe [EUDAMED]
- Model / reference: 11007642 [EUDAMED, FDA GUDID (AccessGUDID)]; MDR\_01\_IIb\_FOR\_ACDV\_Sensis Vibe [EUDAMED]

## Identifiers

- Primary DI / UDI-DI: 04056869010205 [EUDAMED, FDA GUDID (AccessGUDID)]
- Basic UDI-DI: 0405686900144UU [EUDAMED]
- FDA product code: DQK [FDA GUDID (AccessGUDID)]
- EMDN code: Z12050701 [EUDAMED]
- GMDN term: Cardiac catheterization laboratory computer (10980) [FDA GUDID (AccessGUDID)]

## Classification

- Device class: Class IIb [EUDAMED]; Class II [FDA GUDID (AccessGUDID)]
- Regulation number: 870.1425 [FDA GUDID (AccessGUDID)]
- EU registration type: Standard device [EUDAMED]
- EU market of registration: Germany [EUDAMED]
- GUDID distribution status: In Commercial Distribution [FDA GUDID (AccessGUDID)]

## Properties

- Measuring function: Yes [EUDAMED]
- Prescription use: Yes [FDA GUDID (AccessGUDID)]
- Serial number: Yes [FDA GUDID (AccessGUDID)]
- Supplied sterile: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Sterilise before use: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Single use: No [FDA GUDID (AccessGUDID)]
- Implantable: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Reusable: No [EUDAMED]
- Contains latex: No [EUDAMED, FDA GUDID (AccessGUDID)]
- Reprocessed single-use device: No [EUDAMED]
- Contains CMR substances (carcinogenic, mutagenic or toxic to reproduction): No [EUDAMED]
- Incorporates a substance derived from human blood or plasma: No [EUDAMED]
- Contains human tissues or cells: No [EUDAMED]
- Contains animal tissues or cells: No [EUDAMED]
- Contains endocrine-disrupting substances: No [EUDAMED]
- Administering medicine: No [EUDAMED]
- Incorporates a medicinal substance: No [EUDAMED]
- Over-the-counter: No [FDA GUDID (AccessGUDID)]
- Convenience kit: No [FDA GUDID (AccessGUDID)]
- Combination product: No [FDA GUDID (AccessGUDID)]
- HCT/P: No [FDA GUDID (AccessGUDID)]
- Premarket exempt: No [FDA GUDID (AccessGUDID)]
- Lot / batch number: No [FDA GUDID (AccessGUDID)]
- Expiration date: No [FDA GUDID (AccessGUDID)]
- Manufacturing date: No [FDA GUDID (AccessGUDID)]

## Source and citation

- Cite as: Sensis Vibe Combo 11007642, Siemens Healthcare GmbH. deviCERTS medical device catalogue. https://devicerts.com/devices/sensis-vibe-combo-11007642 (data as of 2026-11-28).
- Data sources: [FDA GUDID (AccessGUDID)](https://accessgudid.nlm.nih.gov/), [EUDAMED](https://ec.europa.eu/tools/eudamed), [FDA recalls](https://open.fda.gov/apis/device/recall/)
- Registry records last synced: 2026-11-12
- Statuses computed as of: 2026-11-28
- JSON: <https://devicerts.com/devices/sensis-vibe-combo-11007642.json>
- Licence: [CC BY 4.0](https://creativecommons.org/licenses/by/4.0/). The deviCERTS layout and text are licensed under CC BY 4.0. Registry data remain under their source terms: U.S. FDA data (openFDA, AccessGUDID) are U.S. public domain; EUDAMED data are subject to the European Commission reuse terms.
- Attribution: DevicErts, https://devicerts.com/devices/sensis-vibe-combo-11007642
- Compiled by deviCERTS (devicerts.com) from public medical device registries (EUDAMED, European Commission; U.S. FDA via openFDA; AccessGUDID, U.S. National Library of Medicine). deviCERTS is not an official register and gives no regulatory advice: the record in the registry itself prevails.