Sequoia Spinal System, Model 3300 Series
510(k) K072672
Active
510(k)
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK072672 | Date28 Nov 2007 | StatusActive |
Recalls
-
- Terminated Z-2705-2011
A rounded or partially rounded tip on the screwdriver may not provide the required locking torque to lock the closure top to the polyaxial screw. When the hex tip of the screwdriver is not fully engaged and/or the screwdriver is used off axis, the hex tip of the screwdriver can become rounded or partially rounded.
-
- Terminated Z-1479-2008
Screw Head Dissociation: The polyaxial screw head may dissociate from screw shaft during surgery. Also, the polyaxial screw driver is experiencing difficulty (broken/bent driver shafts, difficulty attaching screws to the driver and jamming of the collet.)
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- NKB
- Device Class
- Class II
- Regulation
- 21 CFR Part 888.3070
- Advisory Committee
- Orthopedic
- FDA Decision
- SESE Substantially Equivalent