Servo Ventilator 300A
510(k) K970839
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Unverified
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)ST | NumberK970839 | Date14 Oct 1997 | StatusActive |
Recalls
-
- Terminated Z-0588-2010
Some Servo Ventilators 300/300A and Servo Ventilator 900C/D/E should not be used with a system that may generate negative pressure below -100 cm H2O (Closed system suctioning) due to the risk of damaging the ventilator's pressure transducers.
-
- Terminated Z-1252-2009
The expiratory valve in the SV 300 ventilator may stay in the open position if the printed circuit board fails.
-
- Terminated Z-0872-2009
SERVO Ventilator 300A could cause hypoventilation if the use of the PC or PRVC mode with the Automode function set to "On" in Neonatal and Pediatric patient ranges with small tidal volumes and short set inspiratory rise time.
-
- Terminated Z-0571-05
Failures of the air and oxygen gas modules occurred because of misuse/non-compliance of directions.
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA Product Code
- CBK
- Device Class
- Class II
- Regulation
- 21 CFR Part 868.5895
- Advisory Committee
- Anesthesiology
- FDA Decision
- ST
Overview
AI-generated summary, 19 Sep 2026
The Servo Ventilator 300A is a continuous-use ventilator designed for facility-based respiratory support. It delivers mechanical ventilation to patients requiring assisted or controlled breathing, with adjustable parameters for pressure, volume, and flow to meet clinical needs in critical care settings.
This device is intended for facility use and is regulated under FDA 510(k) clearance (K970839) as a Class II medical device. It operates on mains power (100V–240V) and requires periodic maintenance, including 1000-hour and 3000-hour overhauls, as outlined in manufacturer service documentation. The Servo Ventilator 300A is not single-use and must be serviced according to specified technical protocols.
Other devices from Siemens Elema AB
Questions
AI-generated answers, 19 Sep 2026
Is the Servo Ventilator 300A designed for use in home settings or only in facility-based environments like hospitals and clinics?
The Servo Ventilator 300A is explicitly designed for facility-based respiratory support only, meaning it is intended for use in hospitals, clinics, or other healthcare settings where continuous mechanical ventilation is required. It is not approved or intended for home use.
How often does the Servo Ventilator 300A require maintenance, and what does that process entail?
The Servo Ventilator 300A requires periodic maintenance, including scheduled overhauls at 1000-hour and 3000-hour intervals. These overhauls must follow the manufacturer’s technical protocols to ensure proper function and safety. Maintenance is not single-use or disposable—it is a reusable device that demands adherence to service guidelines.
What power supply requirements does the Servo Ventilator 300A have, and does it support global electrical standards?
The Servo Ventilator 300A operates on mains power ranging from 100V to 240V, making it compatible with most global electrical standards. However, users must ensure their facility’s power supply matches the device’s voltage and frequency requirements to avoid operational issues.
Is the Servo Ventilator 300A reusable, or is it intended for single-use in a clinical setting?
The Servo Ventilator 300A is not a single-use device. It is designed for continuous, reusable use in facility-based settings, provided it undergoes the required maintenance and overhauls as specified by Siemens Elema AB. Reusability is a key feature for long-term clinical deployment.