Skip to content

Shape-HF Cardiopulmonary Exercise Testing System

by Shape Medical Systems, Inc.

510(k) K090722

Shape-HF Cardiopulmonary Exercise Testing System is a medical device (FDA Class II; FDA product code BTY, Calculator, Predicted Values, Pulmonary Function) made by Shape Medical Systems, Inc., cleared under FDA 510(k) K090722 on 31 Mar 2009, active as of 10 Oct 2026. 2 FDA recalls are linked to it.

Active United States

510(k)

Valid until

No expiry

510(k) clearance is not time-limited

FDA recalls

None ongoing

2 closed

Documents

Trust

Unverified

Source FDA 510(k), synced 10 Oct 2026.

Regulatory status

as of registry sync 10 Oct 2026

RegionAuthorizationNumberDateStatus
United StatesFDA 510(k)SESE Substantially Equivalent · Traditional NumberK090722 Date31 Mar 2009ClearedStatusActive

Official records

From the FDA record

Received by the FDA
19 Mar 2009
FDA decision
31 Mar 2009 (Substantially Equivalent)
FDA review time
12 days from receipt to decision
Public document
510(k) summary on file with the FDA

From the FDA 510(k) record K090722 (openFDA), as read on 10 Oct 2026. Quoted as published, gaps included; nothing on this list is generated.

Recalls

2 recalls · FDA

    • Terminated Z-1060-2015 FDA Class II initiated 23 Dec 2014

    This recall has been initiated because use of the impacted DPIs could result in erroneous testing results potentially leading to incorrect diagnosis and incorrect treatment. The magnitude of error for VO2 or VCO2 parameters could be approximately 20% with use of the impacted product.

    • Terminated Z-2051-2014 FDA Class II initiated 28 May 2014

    Shape Medical has initiated a correction due to a mandatory software upgrade for the Shape HF Cardiopulmonary Testing System prior to use of the impacted DPI lot numbers. Use of the incorrect software version with the impacted DPIs could result in erroneous testing results potentially leading to incorrect diagnosis and incorrect treatment.

Timeline

3 dated events

  1. Cleared FDA 510(k) K090722
  2. Recall initiated FDA recalls Z-2051-2014
  3. Recall initiated FDA recalls Z-1060-2015

Coverage

1 jurisdiction

United States of America
  • Active
  • Expiring
  • Withdrawn
  • Listed
  • Not covered

United States — FDA 510(k) Active

Specification

510(k)

510(k) Number
FDA product code
BTY
Device class
Class II
Regulation
21 CFR Part 868.1890
Advisory committee
Anesthesiology
FDA decision
SESE Substantially Equivalent

Devices cleared under K090722

4 devices · FDA GUDID

Devices that FDA's GUDID lists as cleared or approved under this submission number. One submission usually covers a whole product family, so the sizes and models are listed here once.

Devices cleared under K090722
UDI-DISize and differences
Shape
Shape Medical Systems, Inc.
ShapeB20040011
Single use: yes
On the market
Shape-HF 0008-9001
Shape Medical Systems, Inc.
0008-9001B20089001
Single use: no
On the market
Shape-HF Disposable Patient Interface
Shape Medical Systems, Inc.
Disposable Patient InterfaceB20044001
Single use: yes
On the market
Shape-HF HF001
Shape Medical Systems, Inc.
HF001B200HF001
Single use: no
On the market

Similar devices

Other devices filed under FDA product code BTY (Calculator, Predicted Values, Pulmonary Function).

Other devices from Shape Medical Systems, Inc.

See all devices from Shape Medical Systems, Inc. →