Siemens Enhanced Imaging System
by Siemens Medical Solutions USA, Inc.
510(k) K041166
Siemens Enhanced Imaging System is a Class II device made by Siemens Medical Solutions USA, Inc., cleared under FDA 510(k) K041166 on 17 May 2004, active as of 22 Oct 2026. 10 FDA recalls are linked to it.
Active
Valid until
No expiry
FDA recalls
None ongoingDocuments
Trust
Regulatory status
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| United States | FDA 510(k)SESE | NumberK041166 | Date17 May 2004 | StatusActive |
Official records
- FDA 510(k) database K041166
Recalls
-
- Terminated Z-0061-2017
The detectors on the Symbia S and Symbia T camera systems are supported and moved horizontally and vertically by four linear bearing cars. There is a potential for the linear bearing car to fail allowing the ball bearings to fall out.
-
- Terminated Z-1086-2015
There is a potential for the linear bearing car to fail allowing the ball bearings to fall out.
-
- Terminated Z-0583-2010
- Terminated Z-0584-2010
Some of the hand controllers of the Symbia S & T systems are missing a resistor switch, and the absence of the switch may lead to unintended system motion during camera set up activities.
-
- Terminated Z-1676-2009
- Terminated Z-1677-2009
The pinion gear could crack and cause the gantry to spin freely thereby damaging the Symbia S or T System and potentially injuring a patient during a scan.
-
- Terminated Z-1649-2009
- Terminated Z-1648-2009
Patient injury can occur when the patient is lying on the patient bed and the automatic collimator exchanger is undergoing a collimator change operation at the launch of an acquisition workflow. Detector motions required for this change can cause collision with the patient or pallet if it is in brain scan position, i.e. the patient or pallet is into the field of view.
-
- Terminated Z-0190-2009
- Terminated Z-0189-2009
The protective plastic cap over the CT gantry power switch on the Line Connection Box may come loose, exposing the energized electrical contacts within the switch, thereby causing an electric hazard.
-
- Terminated Z-0548-2007
- Terminated Z-0549-2007
Symbia systems, running on e.soft version 5.5 or earlier, could allow a fully-extended patient pallet to come into contact with the pinhole collimator, resulting in system damage.
Timeline
- Recall initiated FDA recalls Z-0190-2009
- Recall initiated FDA recalls Z-0189-2009
- Recall initiated FDA recalls Z-1649-2009
- Recall initiated FDA recalls Z-1648-2009
- Recall initiated FDA recalls Z-1676-2009
- Recall initiated FDA recalls Z-1677-2009
- Recall initiated FDA recalls Z-0583-2010
- Recall initiated FDA recalls Z-0584-2010
- Recall initiated FDA recalls Z-1086-2015
- Recall initiated FDA recalls Z-0061-2017
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not covered
United States — FDA 510(k) Active
Specification
- 510(k) Number
- FDA product code
- KPS
- Device class
- Class II
- Regulation
- 21 CFR Part 892.1200
- Advisory committee
- Radiology
- FDA decision
- SESE Substantially Equivalent
Similar devices
Other devices filed under FDA product code KPS (System, Tomography, Computed, Emission).
- POSITRON EMISSION COMPUTED TOMOGRAPH SET-1400W by Shimadzu Medical Systems
- QPS (QUANTITATIVE PERFUSION SPECT), QGS (QUANTITATIVE GATED SPECT) by Smv America
- QPS/BPGS/MOCO PROCESSING APPLICATIONS FOR ENTEGRA WORKSTATION by Ge Medical Systems F.I. Haifa
- TRIAD NUCLEAR IMAGING SYSTEM by Trionix Research Laboratory, Inc.
- NEUROQ - PET DP by Syntermed, Inc.
- Discovery NM/CT 670 CZT by GE Healthcare
- CEDARS-SINAI QUANTITATIVE THALLIUM-201 ANALYSIS by Nuccardiac Software, Inc.
- WBR XPRESS PET by Ultraspect, Ltd.
- CEQUAL-QUANTITATIVE ANALYSIS by Nuccardiac Software, Inc.
- VANTAGE 2.1 EXSPECT by Adac Laboratories