Signa Pet/Mr
by GE Medical Systems LLC United States EUDAMED510(k)GUDID
Ref SIGNA PET/MR Product code OUOEMDN Z11020701
Regulatory status
Recall ongoing. 5 ongoing FDA recalls. See recalls
| Region | Authorization | Number | Date | Status |
|---|---|---|---|---|
| European Union | CE/MDR EUDAMED | Number8406821BUG00307H9 | Date25 Jan 2024 | StatusActive |
| United States | GUDID listing GUDID | Number00195278648877 | Date19 Sep 2023 | StatusIn distribution |
| United States | FDA 510(k) 510(k) | NumberK213709 | Date20 Jan 2022 | StatusActive |
Coverage
- Active
- Expiring
- Withdrawn
- Listed
- Not known
- Not covered
- United States
- European Union
Recalls
-
- Ongoing Z-2309-2025
GE HealthCare has become aware that for certain MR systems (SIGNA Artist, SIGNA Voyager, Discovery MR450 1.5T, Optima MR450w 1.5T, SIGNA Architect, SIGNA Premier, SIGNA UHP, Discovery MR750 3.0T, Discovery MR750w 3.0T, SIGNA Hero, SIGNA PET/MR, SIGNA Artist Evo, SIGNA Voyager AIR (China Only), SIGNA Architect AIR (China Only), SIGNA Premier XT (China Only), SIGNA Premier Elite (China Only), SIGNA Premier MAX (China Only), SIGNA Hero XT (China Only), SIGNA Hero Elite (China Only), SIGNA Hero MAX (China Only), SIGNA PET/MR AIR (China Only), SIGNA PET/MR AIR Plus (China Only), the planned maintenance steps to check the functionality of the caster locks on the MR patient table may not have been performed.
-
- Ongoing Z-0520-2025
Gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning.
-
- Ongoing Z-2466-2024
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
-
- Ongoing Z-0595-2022
The magnetic resonance systems could potentially have a cryogen ventilation system that does not meet the venting requirements.
-
- Ongoing Z-2057-2021
GE Healthcare has recently become aware of an issue on the affected products listed below where the "Patient Orientation" button may inadvertently be clicked when intending to click on the "Save RX" button. This changes the prescribed patient orientation on the system prior to running the initial 3-Plane Localizer Scan (See Figure 1 for reference to buttons.) Selecting and saving a patient orientation that does not match the patient's actual position may result in incorrectly annotated and/or flipped images.
-
- Terminated Z-0131-2021
Potential electrical arc flash within the Power Gradient & RF (PGR) Cabinet.
-
- Terminated Z-3029-2017
Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.
-
- Terminated Z-0260-2016
When performing head or neck scans, the currently displayed SAR values could be lower than the actual SAR in the head as predicted by SAR modeling. The predicted HEAD SAR value from the modeling suggests the actual SAR in the head could exceed IEC60601-2-33 limit of 3.2 W/kg for some scans.
Specification
- Trade Name
- Signa Pet/Mr EUDAMEDNa GUDID
- Generic Name
- PET/MRI system
- Model / Reference
- SIGNA PET/MR EUDAMEDSIGNA PET MR GUDID
- Primary DI / UDI-DI
- 00195278648877
- Basic UDI-DI
- 8406821BUG00307H9
- 510(k) Number
- FDA Product Code
- OUO
- EMDN Code
- Z11020701 MRI/pet systems
- GMDN Term
- PET/MRI system
- Device Class
- Class IIa EUDAMEDClass II GUDID
- Regulation Number
- 892.1200
- Market of Registration
- France
- Commercial Distribution
- In Commercial Distribution
Truth per registry
| Field | EUDAMED | GUDID |
|---|---|---|
| Model / Reference | SIGNA PET/MR | SIGNA PET MR |
| Device Class | Class IIa | Class II |
Manufacturer details
- EUDAMED SRN
- US-MF-000001632
- Authorised representative
- GE Medical Systems SCS FR-AR-000000344
Other devices from GE Medical Systems, LLC
Properties & raw data
Show properties, registry sources and raw records
Other properties (21)
- Sterile: No
- Single Use: No
- Implantable: No
- Reusable: No
- Contains Latex: No
- Reprocessed: No
- CMR Substance: No
- Human-derived: No
- Human Tissues: No
- Animal Tissues: No
- Endocrine Disruptor: No
- Administering Medicine: No
- Measuring Function: No
- Incorporates Medicinal Product: No
- Over-the-Counter: No
- Convenience Kit: No
- Combination Product: No
- HCT/P: No
- Premarket Exempt: No
- Lot / Batch Number: No
- Expiration Date: No
Sources (3)
- EUDAMED b8e2fcc0-782f-41c2-b58c-751d591a32df
- FDA UDI 79e9e54c-1e78-45f7-8f03-3b1f1f2bbc8d
- FDA 510(k) K213709